Your browser doesn't support javascript.
loading
Análisis coste-efectividad de dispositivos sanitarios diseñados para prevenir exposiciones percutáneas / Safety-engineered devices to prevent percutaneous injuries: cost-effectiveness analysis on prevention of high-risk exposure
Armadans Gil, Lluís; Fernández Cano, María Isabel; Albero Andrés, Inmaculada; Anglés Mellado, María Luisa; Sánchez García, José María; Campins Martí, Magda; Vaqué Rafart, Josep.
Affiliation
  • Armadans Gil, Lluís; Universitat Autònoma de Barcelona. Hospital Universitari Vall d'Hebron. Barcelona. España
  • Fernández Cano, María Isabel; Universitat Autònoma de Barcelona. Hospital Universitari Vall d'Hebron. Barcelona. España
  • Albero Andrés, Inmaculada; Universitat Autònoma de Barcelona. Hospital Universitari Vall d'Hebron. Barcelona. España
  • Anglés Mellado, María Luisa; Universitat Autònoma de Barcelona. Hospital Universitari Vall d'Hebron. Barcelona. España
  • Sánchez García, José María; Universitat Autònoma de Barcelona. Hospital Universitari Vall d'Hebron. Barcelona. España
  • Campins Martí, Magda; Universitat Autònoma de Barcelona. Hospital Universitari Vall d'Hebron. Barcelona. España
  • Vaqué Rafart, Josep; Universitat Autònoma de Barcelona. Hospital Universitari Vall d'Hebron. Barcelona. España
Gac. sanit. (Barc., Ed. impr.) ; 20(5): 374-381, sept.-oct. 2006. tab
Article in Es | IBECS | ID: ibc-052428
Responsible library: ES1.1
Localization: ES1.1 - BNCS
RESUMEN

Objetivo:

Evaluar la eficiencia de sustituir algunos dispositivos médicos por otros diseñados para prevenir exposiciones percutáneas (EP).

Métodos:

Se han estimado las relaciones coste-efectividad de sustituir algunos dispositivos por otros diseñados para prevenir EP se han considerado los costes de adquisición y de la asistencia a una EP; el número de EP prevenibles por cada dispositivo se ha estimado a partir de las 252 exposiciones ocupacionales notificadas por el personal de un hospital de 1.300 camas de marzo de 2002 a febrero de 2003. Se ha calculado la relación entre el coste adicional de los dispositivos alternativos y el número de EP evitables de alto riesgo (las que cumplían 2 o más de los siguientes EP moderadamente profunda o profunda, por objeto insertado en vaso, o por objeto expuesto a la sangre).

Resultados:

Orden de los dispositivos según la relación coste-efectividad aguja con resguardo para catéter implantado (­2,65 €/EP evitada), seguida por jeringa con resguardo (869,79 €/EP), aguja con aletas con resguardo, equipo de perfusión con válvulas y catéter endovenoso periférico con retracción del fiador. Orden según relación entre coste adicional y número de EP de alto riesgo evitadas aguja para acceso a catéter implantado, seguida por aguja con aletas, jeringa para aguja hipodérmica, catéter endovenoso y equipo de infusión con válvulas.

Conclusiones:

El ahorro en asistencia a EP compensa el coste adicional de algunos dispositivos diseñados para prevenirlas. Un análisis coste-efectividad ayuda a establecer prioridades, pero debe tener en cuenta el riesgo de las EP asociadas con cada dispositivo
ABSTRACT

Objective:

To assess the efficiency of the replacement of several medical devices by engineered sharp injury (SI) prevention devices (ESIPDs).

Methods:

The cost-effectiveness ratios of the replacement of medical devices in use by ESIPDs were estimated their purchasing costs and the direct costs of sharp injury care were taken into account; the number of SI avoidable by each ESIPD was estimated from the 252 occupational SI notified by healthcare workers at a 1,300 bed hospital from March 2002 to February 2003. The relationship between ESIPD additional costs and the number of high-risk SI was estimated (SI were classified as high-risk if they met two or more of the following criteria moderately-deep or deep injury, injury with a device previously inserted in an artery or vein, or with a device exposed to blood).

Results:

ESIPDs order according to cost-effectiveness ratio safety needle for implanted ports (­2.65 €/SI avoided), followed by syringes with protective shield (869.79 €/SI), resheathable winged steel needles, needleless administration sets, and short catheters with protective encasement. ESIPDs order according to relationship between additional costs and number of high-risk sharp injuries avoided safety needles for implanted ports, followed by winged steel needles, hypodermic syringes, short catheter and needleless administration sets.

Conclusions:

Savings in SI care outweigh additional costs of certain ESIPDs. Cost-effectiveness analysis is useful in assigning priorities; however the risks of SI by every device must be taken into account
Subject(s)
Full text: Available Collection: National databases / Spain Health context: SDG3 - Target 3.9 Reduce the amount of deaths produced by dangerous chemicals and the pollution of the air, water and soil / Sustainable Health Agenda for the Americas Health problem: Occupational Exposures / Goal 4: Health financing Database: IBECS Main subject: Personnel, Hospital / Protective Devices / Cost Efficiency Analysis / Occupational Exposure / Infectious Disease Transmission, Patient-to-Professional / Equipment Design / Equipment Safety Type of study: Etiology study / Health economic evaluation / Risk factors Limits: Humans Language: Spanish Journal: Gac. sanit. (Barc., Ed. impr.) Year: 2006 Document type: Article Institution/Affiliation country: Universitat Autònoma de Barcelona/España
Full text: Available Collection: National databases / Spain Health context: SDG3 - Target 3.9 Reduce the amount of deaths produced by dangerous chemicals and the pollution of the air, water and soil / Sustainable Health Agenda for the Americas Health problem: Occupational Exposures / Goal 4: Health financing Database: IBECS Main subject: Personnel, Hospital / Protective Devices / Cost Efficiency Analysis / Occupational Exposure / Infectious Disease Transmission, Patient-to-Professional / Equipment Design / Equipment Safety Type of study: Etiology study / Health economic evaluation / Risk factors Limits: Humans Language: Spanish Journal: Gac. sanit. (Barc., Ed. impr.) Year: 2006 Document type: Article Institution/Affiliation country: Universitat Autònoma de Barcelona/España
...