Comparación de dos sistemas de medida de las concentraciones de glucosa en líquido cefalorraquídeo en la anestesia intradural hiperbara / Comparison of 2 devices to measure spinal fluid glucose concentration in hyperbaric spinal anesthesia
Sánchez Morillo, J; Soliveres Ripoll, J; Estañ Capell, N; Solaz Roldán, C; Viñals Bellido, P; Quiñones Torrelo, C.
Affiliation
  • Sánchez Morillo, J; Hospital Universitario Dr. Peset. Valencia. España
  • Soliveres Ripoll, J; Hospital Universitario Dr. Peset. Valencia. España
  • Estañ Capell, N; Hospital Universitario Dr. Peset. Valencia. España
  • Solaz Roldán, C; Hospital Universitario Dr. Peset. Valencia. España
  • Viñals Bellido, P; Hospital Universitario Dr. Peset. Valencia. España
  • Quiñones Torrelo, C; Hospital Universitario Dr. Peset. Valencia. España
Rev. esp. anestesiol. reanim ; 54(2): 73-77, feb. 2007. ilus, tab
Article in Es | IBECS | ID: ibc-054805
Responsible library: ES1.1
Localization: ES1.1 - BNCS
RESUMEN

OBJETIVOS:

Comparar las mediciones de glucorraquia realizadas en el Laboratorio con el autoanalizador Synchron LX20 PRO(R) (Beckman Coulter) con las realizadas con un glucómetro Ascensia EliteXL&Elite(R) durante la anestesia espinal, después de administrar 10 mg de bupivacaína hiperbara para así disponer de un método rápido y fiable al medir las cifras de glucorraquia durante una anestesia espinal continua y seguir su evolución. PACIENTES Y

MÉTODOS:

Estudio prospectivo de 34 pacientes bajo anestesia espinal continua con un catéter 22 G a través del cual se extrajeron 9 muestras por paciente de líquido cefalorraquídeo para determinar la glucorraquia, la primera antes de la administración del anestésico y los siguientes durante la anestesia espinal hasta el final del bloqueo motor completo. La correlación se midió con el test de Pearson y el grado de concordancia con el método de Bland-Altman.

RESULTADOS:

Se obtuvieron 241 pares de medidas. El coeficiente de correlación fue de r = 0,96 (p < 0,01). La diferencia media entre los dos métodos fue de –1,06 ± 34,82 mg dL–1. El porcentaje de variación (error sistemático) fue de –1,9 ± 11,8%, situándose el intervalo de confianza del 95% entre el –25% y el 21,2%.

CONCLUSIONES:

Se obtiene una buena correlación entre ambos métodos. El error sistemático absoluto y porcentual (bias), es despreciable. Encontrar el 95% de los valores con un margen de un 23% sobre la media del error porcentual nos parece razonable, por lo que consideramos la variación porcentual del error sistemático clínicamente aceptable para poder intercambiar ambos métodos de medida
ABSTRACT

OBJECTIVES:

To compare spinal fluid glucose measurements recorded by the laboratory analyzer Synchron LX20 Pro and the glucometer Ascensia Elite XL during continuous spinal anesthesia after injection of 10 mg of hyperbaric bupivacaine, in order to assess the reliability and speed of the 2 devices for monitoring changes in glucose concentration. PATIENTS AND

METHODS:

Prospective study of 34 patients under continuous spinal anesthesia administered through a 22-gauge catheter; 9 samples of spinal fluid were extracted from each patient for glucose level measurement. The first extraction was before administration of the anesthetic and the remaining ones were during spinal anesthesia until the end of complete motor block. Correlation was assessed with the Pearson test and agreement with the Bland-Altman method.

RESULTS:

A total of 241 pairs of measurements were obtained. The correlation was r=0.96 (P<.01). The mean (SD) difference in measurements from the 2 devices was –1.06 (34.82 mg·dL-1). The percentage of variation (systematic error) was –1.9% (11.8%), placing the 95% confidence interval between -25% and 21.2%.

CONCLUSIONS:

Measurements from the 2 devices are highly correlated. The absolute and percentage systematic error (bias) is negligible. Finding that 95% of measurements are within 23% of the mean seems a fair percentage of error to us. We therefore believe the percentage variation, or systematic error, is clinically acceptable and that either device can be used
Subject(s)
Search on Google
Collection: National databases / Spain Database: IBECS Main subject: Glucose / Anesthesia, Spinal Type of study: Observational study Limits: Aged / Female / Humans / Male Language: Spanish Journal: Rev. esp. anestesiol. reanim Year: 2007 Document type: Article Institution/Affiliation country: Hospital Universitario Dr. Peset/España
Search on Google
Collection: National databases / Spain Database: IBECS Main subject: Glucose / Anesthesia, Spinal Type of study: Observational study Limits: Aged / Female / Humans / Male Language: Spanish Journal: Rev. esp. anestesiol. reanim Year: 2007 Document type: Article Institution/Affiliation country: Hospital Universitario Dr. Peset/España
...