Your browser doesn't support javascript.
loading
Seguridad en la administración intravenosa de medicamentos mediante bombas de infusión inteligentes / Intravenous drug infusion safety through smart pumps
Gómez-Baraza, C; Agustín-Fernández, MJ; Real-Campaña, JM; Abad-Sazatornil, R; Palomo-Jiménez, PI.
Affiliation
  • Gómez-Baraza, C; Hospital Universitario Miguel Servet. Servicio de Farmacia. Zaragoza. España
  • Agustín-Fernández, MJ; Hospital Universitario Miguel Servet. Servicio de Farmacia. Zaragoza. España
  • Real-Campaña, JM; Hospital Universitario Miguel Servet. Servicio de Farmacia. Zaragoza. España
  • Abad-Sazatornil, R; Hospital Universitario Miguel Servet. Servicio de Farmacia. Zaragoza. España
  • Palomo-Jiménez, PI; Hospira Productos Farmacéuticos Hospitalarios, SL. s.c. s.p
Farm. hosp ; 38(4): 276-282, jul.-ago. 2014. ilus, tab
Article in Spanish | IBECS | ID: ibc-131324
Responsible library: ES1.1
Localization: BNCS
RESUMEN

Objetivo:

Analizar el papel de las bombas de infusión inteligentes en la reducción de errores relacionados con la administración de medicación intravenosa.

Método:

Estudio observacional, retrospectivo que analiza la implementación de un sistema de bombas inteligentes de infusión intravenosa (Hospira MedNetTM) y el papel de este sistema de seguridad en la detección de errores en la fase de administración de fármacos, sueros y sangre. Se incluyeron infusiones administradas en los hospitales de día de hematología, oncología, reumatología y oncopediatría. Se analizó la adherencia al sistema de seguridad, el número de errores de programación detectados, los fármacos comúnmente implicados en estos errores y las acciones de mejora.

Resultados:

Durante el periodo de estudio se implementaron 120 bombas inteligentes y se recogieron los datos de 70.028 infusiones. La adherencia al programa de seguridad fue del 62,30% en hematología (6.887 infusiones), del 60,30% en oncología (28.127 infusiones), del 46,50% en reumatología (1.950 infusiones) y del 1,8% en oncopediatría (139 infusiones). Se notificaron 3481 alertas por programación de las bombas fuera de los límites establecidos 2716 de límite relativo y 765 de límite absoluto. En 807 infusiones (2,17%), se evitaron errores que podrían haber tenido consecuencias para los pacientes. Gracias a estos hallazgos, se implementaron una serie de estrategias con objeto de minimizar dichos errores en el futuro.

Conclusiones:

El sistema Hospira MedNetTM intercepta desviaciones con respecto a los protocolos establecidos en la infusión intravenosa, evitando potenciales efectos adversos a pacientes. También permite establecer medidas correctoras e implementar estrategias de mejora (AU)
ABSTRACT

Objective:

To analyze the role of smart infusion pumps in reducing errors related with the administration of intravenous medications.

Method:

Retrospective, observational study analyzing the implementation of a system with smart intravenous infusion pumps (Hospira MedNetTM) and the role of the safety system for the detection of errors during the administration of drugs, sera, and blood. We included infusions administered at the day-care hospitals of hematology, oncology, rheumatology, and oncopediatrics. We analyzed adherence to the safety system, the number of programming errors detected, the commonly implicated drugs in these errors, and improvement actions.

Results:

During the study period, 120 smart pumps were implemented and data on 70,028 infusions were gathered. The rate of adherence to the safety program was 62.30% in hematology (6,887 infusions), 60,30% in oncology (28,127 infusions), 46,50% in rheumatology (1,950 infusions) and 1.8% in oncopediatrics (139 infusions). 3,481 out of the established limits programming alerts were generated by the pumps 2,716 of relative limit and 765 of absolute limit. En 807 infusions (2.17%), errors that could have had consequences for the patients could be prevented. These findings allowed implementing a series of strategies aimed at minimizing these errors in the future.

Conclusions:

The Hospira MedNetTM system detects deviations from the established protocols of intravenous infusion, preventing in this way potential adverse events for the patients. It also allows establishing correction measures and implementing the improvement strategies (AU)
Subject(s)

Full text: Available Collection: National databases / Spain Database: IBECS Main subject: Infusion Pumps / Drug Therapy / Drug-Related Side Effects and Adverse Reactions Type of study: Observational study Limits: Humans Language: Spanish Journal: Farm. hosp Year: 2014 Document type: Article Institution/Affiliation country: Hospira Productos Farmacéuticos Hospitalarios, SL/s.p / Hospital Universitario Miguel Servet/España

Full text: Available Collection: National databases / Spain Database: IBECS Main subject: Infusion Pumps / Drug Therapy / Drug-Related Side Effects and Adverse Reactions Type of study: Observational study Limits: Humans Language: Spanish Journal: Farm. hosp Year: 2014 Document type: Article Institution/Affiliation country: Hospira Productos Farmacéuticos Hospitalarios, SL/s.p / Hospital Universitario Miguel Servet/España
...