Efecto de la dirección de la aguja arterial en la medición del flujo del acceso vascular con métodos dilucionales / Effect of the direction of the arterial needle in the measurement of vascular access flow with dilutional methods
Sánchez Tocino, M Luz; López González, Antonio; Sánchez Martín, Alberto; Villoria González, Silvia; Turrión Cabezas, Vega; Gascón Colorada, María; Burgos Villullas, Marina; Núñez García, M del Pilar; Sánchez Tocino, Hortensia.
Affiliation
  • Sánchez Tocino, M Luz; FRIAT "Las Encinas". Ciudad Rodrigo. España
  • López González, Antonio; Hospital Universitario A Coruña. A Coruña. España
  • Sánchez Martín, Alberto; FRIAT "El Castañar". Béjar. España
  • Villoria González, Silvia; FRIAT "Las Encinas". Ciudad Rodrigo. España
  • Turrión Cabezas, Vega; FRIAT "Los Pinos". Medina del Campo. España
  • Gascón Colorada, María; FRIAT "El Castañar". Béjar. España
  • Burgos Villullas, Marina; FRIAT "Los Pinos". Medina del Campo. España
  • Núñez García, M del Pilar; s.af
  • Sánchez Tocino, Hortensia; Hospital Río Hortega. Valladolid. España
Enferm. nefrol ; 20(4): 305-315, oct.-dic. 2017. graf, tab
Article in Spanish | IBECS | ID: ibc-169611
Responsible library: ES1.1
Localization: BNCS
RESUMEN

Introducción:

La medida del flujo del acceso vascular a través de métodos dilucionales exige la punción retrograda de la aguja arterial. Las últimas recomendaciones de canalización del acceso defienden la punción anterógrada.

Objetivo:

Cuantificar el error de medición producido por la punción anterógrada y analizar su repercusión clínica. Material y

Método:

Estudio transversal multicéntrico. Se analizaron medidas de flujo con Termodilución y Ultrasonografia dilucional en punciones anterógradas y retrogradas. Se utilizó el método Bland-Altman, el coeficiente de correlación intraclase y el índice Kappa para contraste con Eco-Doppler.

Resultados:

Se analizaron 37 fístulas. La punción anterógrada respecto a la retrograda generó medidas que sobrestimaban su diagnóstico en un 65% con Termodilución y 59% con Ultrasonografia Dilucional. El coeficiente de correlación intraclase fue 0,831 para Termodilución y 0,724 para Ultrasonografía Dilucional. El grado de acuerdo clínico entre ambos modos de punción para catalogar a las fístulas con flujo menor de 500ml/min fue del 95% (Índice Kappa=0,841) para Termodilución y del 91% (Índice Kappa=0,760) para Ultrasonografía Dilucional. El grado de acuerdo con Eco-Doppler fue del 86% (Índice Kappa=0.586) y 92% (Índice Kappa=0.720) con Termodilución y Ultrasonografía Dilucional del 78% (Índice Kappa=0.349) y 81% (Índice Kappa=0.400) utilizando la punción anterógrada y retrograda respectivamente.

Conclusión:

La punción anterógrada presenta un sesgo significativo incrementándose en flujos altos. Clínicamente, el grado de concordancia con ambos métodos dilucionales en los distintos tipos de punción y Eco-Dopler es bueno para descarte de intervención (AU)
ABSTRACT

Introduction:

Measurement of vascular access flow through dilutional methods requires retrograde puncture of the arterial needle. The latest cannulation recommendations defend the antegrade puncture.

Aim:

Quantify the measurement error due to the anterograde puncture and analyse its clinical repercussion. Material and

Method:

Multicentric cross-sectional study. Flow measurements were analysed with Termodilution and dilutional Ultrasonography in antegrade and retrograde punctures. We used the Bland-Altman method, the intraclass correlation coefficient and the Kappa index for contrast with Eco-Doppler.

Results:

Thirty-seven fistulas were analysed. The anterograde puncture with respect to the retrograde one generated measures that overestimated the diagnosis by 65% with Thermodilution and 59% with Dilutional Ultrasonography. The intraclass correlation coefficient was 0.831 for Thermodilution and 0.724 for Dilutional Ultrasonography. The degree of clinical agreement between both types of puncture, to classify fistulas with flow less than 500ml/min, was 95% (Kappa Index = 0.841) for Thermodilution and 91% (Kappa Index = 0.760) for Dilutional Ultrasonography. The degree of agreement, according to the antegrade and retrograde puncture (respectively), with Eco-Doppler was 86% (Kappa Index=0.586) and 92% (Kappa Index=0.720); and with Termodilution and Dilutional Ultrasonography of 78% (Kappa Index=0.349) and 81% (Kappa Index=0.400).

Conclusions:

The anterograde puncture presents a significant bias increasing in high flows. Clinically, the degree of agreement with both dilutional methods in different types of puncture and Eco-Doppler indicate a good tool as a discarding intervention (AU)
Subject(s)

Full text: Available Collection: National databases / Spain Database: IBECS Main subject: Thermodilution / Punctures / Ultrasonography, Doppler Type of study: Controlled clinical trial / Practice guideline / Observational study / Prevalence study / Risk factors Limits: Humans Language: Spanish Journal: Enferm. nefrol Year: 2017 Document type: Article Institution/Affiliation country: FRIAT "El Castañar"/España / FRIAT "Las Encinas"/España / FRIAT "Los Pinos"/España / Hospital Río Hortega/España / Hospital Universitario A Coruña/España

Full text: Available Collection: National databases / Spain Database: IBECS Main subject: Thermodilution / Punctures / Ultrasonography, Doppler Type of study: Controlled clinical trial / Practice guideline / Observational study / Prevalence study / Risk factors Limits: Humans Language: Spanish Journal: Enferm. nefrol Year: 2017 Document type: Article Institution/Affiliation country: FRIAT "El Castañar"/España / FRIAT "Las Encinas"/España / FRIAT "Los Pinos"/España / Hospital Río Hortega/España / Hospital Universitario A Coruña/España
...