Your browser doesn't support javascript.
loading
Endoprótesis GORE® EXCLUDER® con rama iliaca para el tratamiento de aneurismas aortoiliacos. Experiencia multicéntrica. Resultados a un año / Treatment of aorto-iliac aneurysms with GORE® EXCLUDER® iliac branch endoprosthesis. Multicentre experience. One year results
Esteban Gracia, J; Maeso Lebrún, J; Dilmé Muñoz, J; Gómez Moya, B; Sancho Serrats, J; Rodríguez Cabeza, P; Cordobés Gual, J.
Affiliation
  • Esteban Gracia, J; Hospital Universitario Germans Trias i Pujol. Servicio de Angiología y Cirugía Vascular. Badalona. España
  • Maeso Lebrún, J; Hospital Vall d'Hebron. Servicio de Angiología y Cirugía Vascular. Barcelona. España
  • Dilmé Muñoz, J; Hospital Santa Creu i Sant Pau. Servicio de Angiología y Cirugía Vascular. Barcelona. España
  • Gómez Moya, B; Hospital Universitari Joan XXIII. Servicio de Angiología y Cirugía Vascular. Tarragona. España
  • Sancho Serrats, J; Clínica Corachan. Servicio de Angiología y cirugía Vascular. Barcelona. España
  • Rodríguez Cabeza, P; Hospital Universitario Josep Trueta. Servicio de Angiología y Cirugía Vascular. Gerona. España
  • Cordobés Gual, J; Hospital Universitario Arnau de Vilanova. Servicio de Angiología y Cirugía Vascular. Lérida. España
Angiol. (Barcelona) ; 70(2): 54-59, mar.-abr. 2018. tab
Article in Spanish | IBECS | ID: ibc-172326
Responsible library: ES1.1
Localization: BNCS
RESUMEN

Introducción:

En el tratamiento endovascular de los aneurismas de aorta abdominal, en el 20-40% de los casos el anclaje en la iliaca no es posible por afectación de la misma. Actualmente es posible el sellado de la endoprótesis en la iliaca externa preservando la permeabilidad de la arteria hipogástrica con dispositivos con ramas iliacas.

Objetivo:

Evaluar la eficacia y la seguridad del dispositivo IBE GORE® y sus resultados a un año para el tratamiento de los aneurismas aortoiliacos con dilatación de la arteria iliaca común. Material y

métodos:

Se trata de un estudio multicéntrico y retrospectivo con un seguimiento de 12 meses. Se incluyen 14 pacientes a los que se les han implantado 18 IBE GORE®. A todos se les ha practicado una TC antes de la cirugía y al menos una TC al mes y otra al año como protocolo de seguimiento.

Resultados:

El éxito técnico ha sido del 100%. A los 30 días, la mortalidad ha sido del 0%. Hubo una trombosis completa de un dispositivo que requirió un bypass cruzado y una oclusión de hipogástrica asintomática. Al año, el diámetro medio de la iliaca ha pasado de 37,3 a 31,4 mm (ha disminuido en el 87,5% de los casos) y no ha habido reintervenciones. La tasa global de endofugas tipo II de la serie ha sido del 21%, la permeabilidad del cuerpo del branch, del 94% (17/18), y la del componente de hipogástrica, del 89% (16/18).

Conclusiones:

El tratamiento de aneurismas aortoiliacos que afectan a la arteria hipogástrica con el dispositivo IBE GORE® es seguro, eficaz y permite preservar la permeabilidad de la hipogástrica (AU)
ABSTRACT

Introduction:

In 20-40% of cases of endovascular aneurysm repair, the regular sealing in common iliac artery is not possible because of the severity of the disease. Nowadays, improved devices provide the possibility of a correct sealing in external iliac artery, preserving the hypogastric patency.

Objective:

The aim of this study is to evaluate the safety and efficacy of the IBE GORE® device and its one year results in the treatment of abdominal aorta-iliac aneurysms. Material and

methods:

A total of 14 patients treated with 18 IBE GORE® devices were evaluated in a multicentre and prospective study, and followed-up during a 12 month period. All of them had a CT scan prior to surgery, at one month, and at one year as required in the follow-up protocol.

Results:

There was 100% technical success. The 30 day follow-up mortality was 0%, with one complete occlusion of the iliac branch that needed a femoro-femoral bypass, and one asymptomatic hypogastric occlusion. The mean iliac artery diameter went from 37.3 to 31.4 mm after one year (decreased in 87.5% of patients), and no re-interventions were needed. The overall type II endoleak was 21%, the patency of the body of the branch was 94% (17/18), and the hypogastric component 89% (16/18).

Conclusions:

Endovascular treatment of aorto-iliac aneurysms affecting the internal iliac artery using the IBE GORE® device is safe, effective, and preserves hypogastric patency (AU)
Subject(s)

Full text: Available Collection: National databases / Spain Database: IBECS Main subject: Stents / Treatment Outcome / Aortic Aneurysm, Abdominal / Endovascular Procedures Type of study: Practice guideline / Observational study Limits: Aged / Humans Language: Spanish Journal: Angiol. (Barcelona) Year: 2018 Document type: Article Institution/Affiliation country: Clínica Corachan/España / Hospital Santa Creu i Sant Pau/España / Hospital Universitari Joan XXIII/España / Hospital Universitario Arnau de Vilanova/España / Hospital Universitario Germans Trias i Pujol/España / Hospital Universitario Josep Trueta/España / Hospital Vall d'Hebron/España

Full text: Available Collection: National databases / Spain Database: IBECS Main subject: Stents / Treatment Outcome / Aortic Aneurysm, Abdominal / Endovascular Procedures Type of study: Practice guideline / Observational study Limits: Aged / Humans Language: Spanish Journal: Angiol. (Barcelona) Year: 2018 Document type: Article Institution/Affiliation country: Clínica Corachan/España / Hospital Santa Creu i Sant Pau/España / Hospital Universitari Joan XXIII/España / Hospital Universitario Arnau de Vilanova/España / Hospital Universitario Germans Trias i Pujol/España / Hospital Universitario Josep Trueta/España / Hospital Vall d'Hebron/España
...