Infrasensado auricular intermitente de marcapasos VDD en pacientes con repolarización sensible a bradicardia: un potencial mecanismo de arritmias ventriculares / Intermittent atrial undersensing in single-lead VDD pacemakers in patients with bradycardia-sensitive repolarization: a possible mechanism for ventricular arrhythmia
Rev. esp. cardiol. (Ed. impr.)
; 63(2): 229-232, feb. 2010. ilus
Article
in Spanish
| IBECS
| ID: ibc-76237
Responsible library:
ES1.1
Localization: BNCS
RESUMEN
Es conocida la escasa repercusión clínica que tiene el fallo de sensado auricular intermitente en portadores de marcapasos VDD. No obstante, en pacientes con un sustrato de repolarización alterada mediada por bradicardia, este fallo puede generar oscilaciones RR ocasionadas por los cambios de modo, que pueden ser el desencadenante para iniciar una torsade de pointes. Exponemos nuestros hallazgos con 3 casos que ilustran este mecanismo desencadenante de arritmias ventriculares (AU)
ABSTRACT
Intermittent atrial undersensing in patients with single-lead VDD pacemakers is regarded as being of limited clinical significance. Nevertheless, in patients with bradycardia-mediated repolarization abnormalities, undersensing could result in RR-interval oscillations due to changes in pacemaker mode and this could act as a trigger for the initiation of torsades de pointes. We report our findings in three patients, which demonstrate the functioning of this trigger mechanism for ventricular arrhythmias (AU)
Search on Google
Collection:
National databases
/
Spain
Database:
IBECS
Main subject:
Pacemaker, Artificial
/
Arrhythmias, Cardiac
/
Bradycardia
Type of study:
Diagnostic study
Limits:
Female
/
Humans
/
Male
Language:
Spanish
Journal:
Rev. esp. cardiol. (Ed. impr.)
Year:
2010
Document type:
Article
Institution/Affiliation country:
Consorcio Hospital General Universitario de Valencia/España
/
Hospital General de Castellón/España
/
Instituto Cardiovascular del Consorcio Hospital General Universitario de Valencia/España