High impedance and low energy stimulation with bipolar ultrasmall electrode surface area steroid eluting lead (CapSure Z)
REBLAMPA Rev. bras. latinoam. marcapasso arritmia ; 8(n.esp): 214-8, out. 1995. graf
Article in English | LILACS | ID: lil-165654
Responsible library: BR44.1
ABSTRACT
Thirty-five patients received a Capsure Z Medtronic bipolar model 5034 for the ventricle and 5534 for the atrium. These have 1.2 mm2 totally proprous, platinized, steroid eluting distal electrodes. The control group was repesented by 14 consective patients reciving CapSure SP medtronic model 5024 for the ventricle and 5524 for the atrium. At implant and during the follow-up period (6 months), CapSure Z and SP showed similar low pacing theresholds without early peaking in atrium (0.5 +/- 0.26 V vs 0 +/- 0.21 V at 0.5 msec P=NS., 0.06 +/- 0.01 msec at 1.6 V P=NS, respectively) and while Capsure Z showed a lewr value at six months in ventricle (0.3 +/- 0.1 V vs ).4 +/- 0.23 V P=NS, respectively) and while Capsure Z showed a lower value at six months in ventricle (0.3 +/- V vs 0.4 +/- 0.23 V P=NS at 0.5 msec, 0.07 +/- 0.03 msec. vs 0.1 +/- 0.02 msec. P<0.02 at 1.6V, repectively). P and R wave sensing of CapSure Z was better thant that of CapSure SP at implant (P wave + 5 +/- 2.5 mV vs 3.80 +/- 1.95 m;R wave = 15.2 +/- 6.4 mV vs 13.13 +/- 5.5 mV, respecitively) and during the follow-up period achieving statistical significance at the 6 th mont oly for P wave (P wave = 3.33 +/- 1.6 mV vs 2.61 +/- 1.05 mV P<0.05; R wave = 13.9 +/- 5.17 vs 10.8 +/- 5.75 mV, P=NS). CapSure Z atrial and ventricular pacing impedance were double than that of CapSure SP one at implant (atrium 1050 +/- 214 vs 491 +/- 51; ventricle 1296 +/- 236 vs 481 +/- 81, p< 0.0001) and during the follow-up period (atrium 1081 +/- 185vs553 +/- 60; ventricle 1186 +/- 256 vs 656+/- 68, P< 0.0001).
Subject(s)
Search on Google
Collection: International databases Database: LILACS Main subject: Pacemaker, Artificial / Cardiac Pacing, Artificial / Electric Conductivity / Electrodes / Longevity Language: English Journal: REBLAMPA Rev. bras. latinoam. marcapasso arritmia Journal subject: Cardiology Year: 1995 Document type: Article
Search on Google
Collection: International databases Database: LILACS Main subject: Pacemaker, Artificial / Cardiac Pacing, Artificial / Electric Conductivity / Electrodes / Longevity Language: English Journal: REBLAMPA Rev. bras. latinoam. marcapasso arritmia Journal subject: Cardiology Year: 1995 Document type: Article
...