Your browser doesn't support javascript.
loading
Impacto de um programa de avaliação da qualidade da imagem nos serviços de mamografia do Distrito Federal / Impact of a program of diagnostic imaging quality control in mammography centers of the Federal District, Brazil
Corrêa, Rosangela da Silveira; Peixoto, João Emílio; Silver, Lynn Dee; Dias, Cintia Melazo; Nogueira, Maria do Socorro; Hwang, Suy Ferreira; Ferreira, Rubemar de Souza.
Affiliation
  • Corrêa, Rosangela da Silveira; Comissão Nacional de Energia Nuclear. Centro Regional de Ciências Nucleares do Centro-Oeste. Abadia de Goiás. BR
  • Peixoto, João Emílio; Universidade Federal do Rio de Janeiro. Rio de Janeiro. BR
  • Silver, Lynn Dee; Departamento de Saúde de Nova York. Nova York. US
  • Dias, Cintia Melazo; Comissão Nacional de Energia. Distrito do Planalto Central. Brasília. BR
  • Nogueira, Maria do Socorro; Comissão Nacional de Energia. Centro de Desenvolvimento da Tecnologia Nuclear. Belo Horizonte. BR
  • Hwang, Suy Ferreira; Comissão Nacional de Energia. Centro Regional de Ciências Nucleares do Nordeste. Recife. BR
  • Ferreira, Rubemar de Souza; Comissão Nacional de Energia Nuclear. Centro Regional de Ciências Nucleares do Centro-Oeste. Abadia de Goiás. BR
Radiol. bras ; Radiol. bras;41(2): 109-114, mar.-abr. 2008. graf
Article in Pt | LILACS | ID: lil-482997
Responsible library: BR734.1
RESUMO

OBJETIVO:

Esta pesquisa visou a avaliar a qualidade dos serviços de mamografia do Distrito Federal e o impacto de uma intervenção de inspeção e capacitação.

MATERIAIS E MÉTODOS:

Foram estudados 41 serviços de mamografia no Distrito Federal no período de 2000 a 2002. A intervenção consistiu na inspeção inicial seguida de um treinamento e notificação oficial da Vigilância Sanitária. Os resultados de qualidade da imagem foram comparados "antes" e "depois" da intervenção.

RESULTADOS:

O estudo demonstrou que dos 36 serviços que completaram a pesquisa, nenhum estava acima de 90 por cento de conformidade antes da intervenção. Após a intervenção, dez unidades atingiram mais de 90 por cento. As principais melhorias foram em relação aos chassis, compressão da mama e visualização de microcalcificações.

CONCLUSÃO:

Apesar de o Distrito Federal dispor de muitos serviços, na sua maioria não eram de qualidade. A intervenção foi eficaz para a melhoria da qualidade, porém, torna-se necessária uma ação continuada para resolver os problemas restantes e aumentar o impacto.
ABSTRACT

OBJECTIVE:

The present study was aimed at evaluating the quality and the impact of an intervention involving inspection and education in mammography centers of the Federal District, Brazil. MATERIALS AND

METHODS:

Forty one mammography centers in the Federal District were studied in the period between 2000 and 2002. The intervention involved an initial inspection followed by a training activity and notification of mammography centers by the Federal District Sanitary Vigilance authority. The imaging quality was compared before and after the intervention.

RESULTS:

None of the 36 centers which completed the study reached more than 90 percent compliance with the standard imaging quality prior to the interventions, whereas ten were above 90 percent afterwards. Major improvements were observed in chassis maintenance, breast compression and visualization of microcalcifications.

CONCLUSION:

Despite the availability of a great number of mammography centers in the Federal District, most of them did not meet the required quality standards. The intervention has shown to be effective for improving the imaging quality, however a continued action is required to solve the remaining problems and increase the impact of the program.
Subject(s)
Key words
Full text: 1 Collection: 01-internacional Database: LILACS Main subject: Quality Control / Mammography / Technology, Radiologic / Evaluation Studies as Topic Type of study: Diagnostic_studies / Evaluation_studies Country/Region as subject: America do sul / Brasil Language: Pt Journal: Radiol. bras / Radiol. bras. (Online) / Radiologia brasileira (Impresso) Journal subject: DIAGNOSTICO POR IMAGEM / RADIOLOGIA Year: 2008 Document type: Article Affiliation country: Brazil / United States Country of publication: Brazil
Full text: 1 Collection: 01-internacional Database: LILACS Main subject: Quality Control / Mammography / Technology, Radiologic / Evaluation Studies as Topic Type of study: Diagnostic_studies / Evaluation_studies Country/Region as subject: America do sul / Brasil Language: Pt Journal: Radiol. bras / Radiol. bras. (Online) / Radiologia brasileira (Impresso) Journal subject: DIAGNOSTICO POR IMAGEM / RADIOLOGIA Year: 2008 Document type: Article Affiliation country: Brazil / United States Country of publication: Brazil