Calibração da creatinina sérica com uso de equação de regressão: influência na performance da fórmula do estudo MDRD / Serum creatinine calibration with regression equation: influence on MDRD formula performance
Rev. HCPA & Fac. Med. Univ. Fed. Rio Gd. do Sul
; 29(3): 188-193, 2009. ilus
Article
in Pt
| LILACS
| ID: lil-552650
Responsible library:
BR18.1
RESUMO
Diferenças nos métodos de medida da creatinina sérica podem determinar amplas variações na taxa de filtração glomerular (TFG) estimada com fórmulas. Para a padronização da medida da creatinina, deve ser usado método calibrado rastreável para medida de referência com ID-MS (isotope dilution mass spectrometry). Objetivo:
Avaliar a TFG estimada com a equação MDRD (Modification of Diet in Renal Disease) original (MDRDo, creatinina método não calibrado) e a equação MDRD re-expressa (MDRDr, método calibrado por ID-MS), comparando-as com a TFG medida pelo 51Cr-EDTA (método padrão) em indivíduos normais.Métodos:
Foram avaliados 101 indivíduos, com idade média de 38±12 anos, sendo 45 homens. A TFG foi medida pela técnica de injeção única do 51Cr-EDTA (TFG 51Cr-EDTA) e estimada pelas equações MDRDo 186 x creatinina sérica-1,154 x idade-0,203 x 0,742 (se mulher) x 1,210 (se negro) e MDRDr, substituindo-se o valor 186 por 175 na equação. A creatinina sérica foi medida pelo método de Jaffe não calibrado e transformada em calibrado com a fórmula y=1,07x-0,249, obtida previamente por regressão. A concordância entre os métodos foi avaliada através da análise de Bland&Altman.Resultados:
Os valores médios para as TFG 51Cr-EDTA, MDRDr e MDRDo foram de 105±18, 102±21 e 84±13 ml/min/1,73 m², respectivamente. Acurácia (percentual de casos de TFG estimada que não desviam em mais de 15% do valor medido) foi maior com o uso da MDRDr em relação à MDRDo (57% vs. 35%, P=0,002). O viés (diferença entre TFG medida e estimada) para 51Cr-EDTA e MDRDr foi de 3±23 ml/min/1,73 m². Para 51Cr-EDTA e MDRDo o viés foi significativamente maior, sendo de 21±18 ml/min/1,73 m². No entanto, a precisão, avaliada como desvio padrão do viés indicou elevada dispersão nos dois casos.Conclusão:
O uso da equação re-expressa do MDRD, empregando a creatinina calibrada, produz uma estimativa mais acurada da TFG do que a equação original do MDRD.ABSTRACT
Differences in methods of measurement of serum creatinine may provide wide variations in glomerular filtration rate (GFR) estimated with formulas. To standardize the measurement of creatinine, calibrated methods should be used, traceable to the reference ID-MS (isotope dilution mass spectrometry) method. Aim:
To evaluate the performance of GFRs estimated with the original Modification of Diet in Renal Disease study equation (MDRDo; non-calibrated creatinine method) and with the re-expressed MDRD equation (MDRDr; ID-MS creatinine), comparing them with the GFR measured by 51Cr-EDTA (standard method ) in normal adults.Methods:
101 subjects, aged 38±12 years, 45 (45%) men were evaluated. GFR was measured by single-injection 51Cr-EDTA (GFR 51Cr-EDTA) technique and estimated by the following equations - MDRDo 186 x serum creatinine-1. 154 x age-0.203 x 0.742 (if female) x 1.210 (if black), and MDRDr, replacing the value 186 by 175 in the equation. Serum creatinine was measured by a non-calibrated Jaffes method and transformed into calibrated with the formula y=1.07x-0.249, previously obtained by regression. The agreement between methods was assessed by the Bland&Altman analyses.Results:
The mean GFR 51Cr-EDTA, MDRDr and MDRDo were 105±18, 102±21 and 84±13 ml/min/1, 73 m2, respectively. There was no agreement between 51Cr-EDTA and MDRDo GFR (P <0.001), but it was present between 51Cr-EDTA and MDRDr GFR (P=0.149). Accuracy (percentage of cases of estimated GFR within 15% of measured value) was higher with the use of MDRDr in comparison to MDRDo (57% vs. 35%, P=0.002). Bias (diference between measured and estimated GFR) for 51Cr-EDTA and MDRDr was 3±23 ml/min/1.73 m². For 51Cr-EDTA and MDRDo the bias was significantly higher, 21±18 ml/min/1.73 m². However, the precision, evaluated as standard deviation of bias indicated a huge variation in both cases.Conclusion:
The re-expressed MDRD equation, using calibrated creatinine, is a more accurate estimation of GFR than.Key words
Full text:
1
Collection:
01-internacional
Database:
LILACS
Main subject:
Creatinine
/
Renal Insufficiency, Chronic
/
Glomerular Filtration Rate
Type of study:
Evaluation_studies
/
Prognostic_studies
Limits:
Adult
/
Humans
/
Male
Language:
Pt
Journal:
Rev. HCPA & Fac. Med. Univ. Fed. Rio Gd. do Sul
Journal subject:
MEDICINA
Year:
2009
Document type:
Article
Affiliation country:
Brazil
Country of publication:
Brazil