Personalización de las fórmulas de cálculo de lente intraocular / Personalization of the intraocular lens calculation formulas
Hernández Silva, Juan R.; Padilla González, Carmen M.; Ramos López, Meisy; Ríos Cazo, Reinaldo; Río Torres, Marcelino.
Affiliation
  • Hernández Silva, Juan R.; Hospital Oftalmológico Docente Ramón Pando Ferrer. Ciudad de La Habana. CU
  • Padilla González, Carmen M.; Hospital Oftalmológico Docente Ramón Pando Ferrer. Ciudad de La Habana. CU
  • Ramos López, Meisy; Hospital Oftalmológico Docente Ramón Pando Ferrer. Ciudad de La Habana. CU
  • Ríos Cazo, Reinaldo; Hospital Oftalmológico Docente Ramón Pando Ferrer. Ciudad de La Habana. CU
  • Río Torres, Marcelino; Hospital Oftalmológico Docente Ramón Pando Ferrer. Ciudad de La Habana. CU
Rev. cuba. oftalmol ; 17(2)jul.-dic. 2004. ilus, tab
Article in Spanish | LILACS | ID: lil-629435
Responsible library: CU1.1
RESUMEN
Este trabajo tiene la finalidad de evaluar los resultados refractivos que se alcanzan con la introducción de las constantes personalizadas en las fórmulas de tercera generación para el cálculo del lente intraocular. Se realizó un estudio descriptivo prospectivo cuyo universo está constituido por los pacientes con catarata senil y se seleccionó una muestra aleatoria de 597 casos. Los pacientes fueron clasificados en 7 grupos según las mediciones preoperatorios del promedio queratométrico y la longitud axial. Para dar salida al objetivo propuesto fueron utilizadas las siguientes variables longitud axial, promedio queratométrico, componente esférico esperado (sep), componente esférico obtenido (ceo), dioptría de lente intraocular implantado (<16, de 17-19, 20-23, 24-27, 28-30) y resultados refractivos. El análisis estadístico de los resultados se realizó, mediante un análisis de varianza, la Prueba T de comparación de medias para datos pareados y Ji cuadrado. Para todo lo anterior se utilizó un nivel de confiabilidad del 95 %. La utilización de las constantes personalizadas en el cálculo de los lentes intraoculares, incrementó en más de un 20 % los buenos resultados refractivos, sin que existan diferencias en los diferentes grupos de pacientes, el tipo de fórmula utilizada y el poder dióptrico del lente. Los modelos de lente S3652 y S3602 presentaron un porcentaje elevado de hipocorrección. La agudeza visual alcanzada en todos los casos fue superior a 0.5 resultado comparable con los estudios internacionales recientes. La profundidad de la cámara anterior estimada preoperatoriamente se correspondió con el valor real posoperatorio, lo que habla de una posición efectiva del lente adecuada.
ABSTRACT
This paper is aimed at evaluating the refractive results attained with the introduction of personalized constants in the third generation formulas for the intraocular lens calculation. A descriptive and prospective study was conducted among the patients with senile cataract, and a randomized sample of 597 cases was selected. Patients were classified into 7 groups according to the preoperative measurements of the keratometric average and the axial longitude. To obtain the proposed objective, the following variables were used axial longitude, keratometric average, expected spherical component (sep), obtained spherical component (ceo), dioptry of implanted ocular lens (<16, of 17-19, 20-23, 24-27, 28-30) and refractive results. The statistical analysis of the results was made by a varianze analysis, the T test of comparison of means for matched data and Ji square. A confidence level of 95.5 was used. The utilization of personalized constants in the calculation of intraocular lens increased by more than 20 % the good refractive results. No differences were observed in the diverse groups of patients, the type of formula used and the dioptric power of the lens.The S3652 and S3602 lens models presented a high percentage of hypocorrection. The visual acuity attained in all cases was over 0.5, a result that is comparable with that of recent international studies. The preoperatively estimated corresponded to the real postoperative value, which evidences an adequate effective position of the lens.

Full text: Available Collection: International databases Database: LILACS Type of study: Controlled clinical trial / Observational study / Prognostic study Language: Spanish Journal: Rev. cuba. oftalmol Journal subject: Ophthalmology Year: 2004 Document type: Article Affiliation country: Cuba Institution/Affiliation country: Hospital Oftalmológico Docente Ramón Pando Ferrer/CU
Full text: Available Collection: International databases Database: LILACS Type of study: Controlled clinical trial / Observational study / Prognostic study Language: Spanish Journal: Rev. cuba. oftalmol Journal subject: Ophthalmology Year: 2004 Document type: Article Affiliation country: Cuba Institution/Affiliation country: Hospital Oftalmológico Docente Ramón Pando Ferrer/CU
...