Accuracy of five electronic foramen locators with different operating systems: an ex vivo study
Journal of Applied Oral Science; Vasconcelos, Bruno Carvalho de; Bueno, Michelli de Medeiros; Luna-Cruz, Suyane Maria; Duarte, Marco Antonio Hungaro; Fernandes, Carlos Augusto de Oliveira.
Affiliation
  • Vasconcelos, Bruno Carvalho de; Federal University of Ceará. Sobral School of Dentistry.
  • Bueno, Michelli de Medeiros; Federal University of Ceará. Sobral School of Dentistry.
  • Luna-Cruz, Suyane Maria; Federal University of Ceará. Sobral School of Dentistry.
  • Duarte, Marco Antonio Hungaro; Federal University of Ceará. Sobral School of Dentistry.
  • Fernandes, Carlos Augusto de Oliveira; Federal University of Ceará. Sobral School of Dentistry.
J. appl. oral sci ; J. appl. oral sci;21(2): 132-137, Mar-Apr/2013. tab
Article in En | LILACS | ID: lil-674356
Responsible library: BR1.1
ABSTRACT

Objective:

The aim of this study was to evaluate, ex vivo, the precision of five electronic root canal length measurement devices (ERCLMDs) with different operating systems the Root ZX, Mini Apex Locator, Propex II, iPex, and RomiApex A-15, and the possible influence of the positioning of the instrument tips short of the apical foramen. Material and

Methods:

Forty-two mandibular bicuspids had their real canal lengths (RL) previously determined. Electronic measurements were performed 1.0 mm short of the apical foramen (-1.0), followed by measurements at the apical foramen (0.0). The data resulting from the comparison of the ERCLMD measurements and the RL were evaluated by the Wilcoxon and Friedman tests at a significance level of 5%.

Results:

Considering the measurements performed at 0.0 and -1.0, the precision rates for the ERCLMDs were 73.5% and 47.1% (Root ZX), 73.5% and 55.9% (Mini Apex Locator), 67.6% and 41.1% (Propex II), 61.7% and 44.1% (iPex), and 79.4% and 44.1% (RomiApex A-15), respectively, considering ±0.5 mm of tolerance. Regarding the mean discrepancies, no differences were observed at 0.0; however, in the measurements at -1.0, the iPex, a multi-frequency ERCLMD, had significantly more discrepant readings short of the apical foramen than the other devices, except for the Propex II, which had intermediate results. When the ERCLMDs measurements at -1.0 were compared with those at 0.0, the Propex II, iPex and RomiApex A-15 presented significantly higher discrepancies in their readings.

Conclusions:

Under the conditions of the present study, all the ERCLMDs provided acceptable measurements at the 0.0 position. However, at the -1.0 position, the ERCLMDs had a lower precision, with statistically significant differences for the Propex II, iPex, and RomiApex A-15.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: LILACS Main subject: Root Canal Preparation / Tooth Apex / Dental Instruments / Dental Pulp Cavity / Odontometry Type of study: Evaluation_studies Limits: Humans Language: En Journal: J. appl. oral sci Journal subject: ODONTOLOGIA Year: 2013 Document type: Article Affiliation country: Brazil Country of publication: Brazil

Full text: 1 Collection: 01-internacional Database: LILACS Main subject: Root Canal Preparation / Tooth Apex / Dental Instruments / Dental Pulp Cavity / Odontometry Type of study: Evaluation_studies Limits: Humans Language: En Journal: J. appl. oral sci Journal subject: ODONTOLOGIA Year: 2013 Document type: Article Affiliation country: Brazil Country of publication: Brazil