Your browser doesn't support javascript.
loading
Cierre percutáneo de la comunicación interauricular con el dispositivo Amplatzer®: 15 años de experiencia / Percutaneous closure of atrial septal defects with the Amplatzer® device: 15 years of experience
Zabal-Cerdeira, Carlos; García-Montes, José Antonio; Sandoval-Jones, Juan Pablo; Calderón-Colmenero, Juan; Patiño-Bahena, Emilia; Juanico-Enríquez, Antonio; Buendía-Hernández, Alfonso.
Affiliation
  • Zabal-Cerdeira, Carlos; Instituto Nacional de Cardiología Ignacio Chávez. Departamento de Hemodinámica. México. MX
  • García-Montes, José Antonio; Instituto Nacional de Cardiología Ignacio Chávez. Departamento de Hemodinámica. México. MX
  • Sandoval-Jones, Juan Pablo; Instituto Nacional de Cardiología Ignacio Chávez. Departamento de Hemodinámica. México. MX
  • Calderón-Colmenero, Juan; Instituto Nacional de Cardiología Ignacio Chávez. Departamento de Hemodinámica. México. MX
  • Patiño-Bahena, Emilia; Instituto Nacional de Cardiología Ignacio Chávez. Departamento de Hemodinámica. México. MX
  • Juanico-Enríquez, Antonio; Instituto Nacional de Cardiología Ignacio Chávez. Departamento de Hemodinámica. México. MX
  • Buendía-Hernández, Alfonso; Instituto Nacional de Cardiología Ignacio Chávez. Departamento de Hemodinámica. México. MX
Arch. cardiol. Méx ; 84(4): 250-255, oct.-dic. 2014. ilus
Article in Spanish | LILACS | ID: lil-744058
Responsible library: MX1.1
RESUMEN

Introducción:

El cierre percutáneo de la comunicación interauricular es una alternativa al tratamiento quirúrgico.

Objetivos:

Comunicar los resultados del cierre percutáneo de la comunicación interauricular con el dispositivo Amplatzer®.

Método:

Se incluyeron los pacientes sometidos a hemodinamia de septiembre de 1997 a diciembre de 2011.

Resultados:

Se procedió al cierre del defecto en 721 pacientes y en 85 no se cerró por considerarse no apto anatómicamente. El diámetro expandido del defecto fue de 23 ± 6.7 mm (límites, 5-42). Todos los dispositivos se colocaron con éxito. En 15 casos (2.1%) se cambió el dispositivo por considerarse inestable y en 6 (0.8%) se presentó embolización en las primeras 24 h, 4 fueron rescatados y recolocados (éxito final con intención de tratamiento 719/806-89.2%). El control inmediato mostró oclusión completa en 247 pacientes (34.3%), fuga intradispositivo en 395 (54.9%), ligera en 75 (10.4%) y moderada en 2 (0.3%). Se hace seguimiento en 626 pacientes (87.1%) durante 33 ± 27.5 meses, presentándose cefalea en 54 (8.6%), arritmias supraventriculares nuevas en 10 (1.6%), complicaciones en el sitio de punción en 4 (0.6%) y accidente vascular cerebral en uno (0.1%).

Conclusión:

El diseño del dispositivo Amplatzer® permite el cierre de la comunicación interauricular con una técnica sencilla y con seguridad para el paciente.
ABSTRACT

Introduction:

Percutaneous closure of atrial septal defects is an alternative to surgical treatment.

Objectives:

We report the results of percutaneous closure of atrial septal defects with the Amplatzer® device.

Method:

We include patients taken to the catheterization laboratory from September 1997 to December 2011.

Results:

We proceeded with defect closure in 721 patients and in 85 the defect was considered not suitable for closure. The stretched diameter of the defect was 23 ± 6.7 mm (limits, 5-42). All devices were positioned successfully. In 15 cases (2.1%) the device was changed due to instability and in 6 (0.8%) the device embolized within the first 24 h, 4 were retrieved and repositioned (final success with intention to treat 719/806-89.2%). Immediate control showed complete closure in 247 patients (34.3%), leak through the device in 395 (54.9%), mild residual leak in 75 (10.4%) and moderate in 2 (0.3%). We have follow-up in 626 patients (87.1%) for 33 ± 27.5 months, with headache in 54 (8.6%), new supraventricular arrhythmia in 10 (1.6%), puncture site complications in 4 (0.6%), and stroke in one (0.1%).

Conclusion:

The design of the Amplatzer® device allows effective closure of atrial septal defects with a simple technique and patient safety.
Subject(s)


Full text: Available Collection: International databases Database: LILACS Main subject: Septal Occluder Device / Heart Septal Defects, Atrial Limits: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Infant / Male Language: Spanish Journal: Arch. cardiol. Méx Journal subject: Cardiology Year: 2014 Document type: Article Affiliation country: Mexico Institution/Affiliation country: Instituto Nacional de Cardiología Ignacio Chávez/MX

Full text: Available Collection: International databases Database: LILACS Main subject: Septal Occluder Device / Heart Septal Defects, Atrial Limits: Adolescent / Adult / Aged / Child / Child, preschool / Female / Humans / Infant / Male Language: Spanish Journal: Arch. cardiol. Méx Journal subject: Cardiology Year: 2014 Document type: Article Affiliation country: Mexico Institution/Affiliation country: Instituto Nacional de Cardiología Ignacio Chávez/MX
...