Your browser doesn't support javascript.
loading
Dual-phase gas-permeation flow-injection thermometric analysis for the determination of carbon dioxide.
Liu, S J; Tubino, M.
Affiliation
  • Liu SJ; Instituto de Quimica, Universidade Estadual de Campinas (UNICAMP), Caixa Postal 6154, CEP 13083-970, Campinas, SP, Brazil.
Talanta ; 47(3): 711-7, 1998 Nov.
Article in En | MEDLINE | ID: mdl-18967375
A flow-injection configuration based on a dual-phase gas-permeation system from a liquid donor to a gas acceptor stream with a thermistor flow-through detector is proposed for the direct analysis of the gas in the acceptor. This system was applied for the determination of carbon dioxide (in the form of carbonate) using the following chemical reaction: CO(2)(g)+2NH(3)(g)+H(2)O(g)=(NH(4))(2)CO(3)(s), with a linear response from 1x10(-3) to 50x10(-3) mol l(-1) of CO(3)(2-). Carbon dioxide was produced in the liquid donor and permeated into the gaseous acceptor stream of air/water vapor. The detection limit is 1x10(-3) mol l(-1) of carbonate, and a sampling frequency of 60 h(-1) is achieved with a relative standard deviation of 4.1% for replicate injections. The dual-phase gas-permeation flow-injection manifold, along with the membrane and phase separations, as well as the chemical reaction, provides enhanced selectivity when compared with the system employing a liquid acceptor stream, as serious interferents in this system, for instance, acetate and formate, among others, do not interfere in the proposed system.
Search on Google
Collection: 01-internacional Database: MEDLINE Language: En Journal: Talanta Year: 1998 Document type: Article Affiliation country: Brazil Country of publication: Netherlands
Search on Google
Collection: 01-internacional Database: MEDLINE Language: En Journal: Talanta Year: 1998 Document type: Article Affiliation country: Brazil Country of publication: Netherlands