Characterization of a Si(Li) Detector for PIXE Analysis.
Rodríguez-Fernández, L; Miranda, J; Oliver, A.
Affiliation
  • Rodríguez-Fernández L; Instituto de Física, Universidad Nacional Autónoma de México, Apartado Postal 20-364, México, D.F. 01000, Mexico.
J Xray Sci Technol ; 4(3): 221-46, 1994 Jan 01.
Article in En | MEDLINE | ID: mdl-21307495
The characterization of a Si(Li) detector used for PIXE analysis is presented. The main detector parameters are indicated, and the different methods of determining them are examined. Also, the detection efficiency has been measured in the 1.4-100 keV photon energy range, using calibrated radioactive sources and PIXE, to obtain and compare the fitted parameters. Finally, the fit of an analytic function to the measured efficiency values and the efficiency in parametric form are compared, and the advantages observed for each are noted.

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Xray Sci Technol Journal subject: RADIOLOGIA Year: 1994 Document type: Article Affiliation country: Mexico Country of publication: Netherlands