Fast frequency sweeping in resonance-tracking SPM for high-resolution AFAM and PFM imaging.
Nanotechnology
; 23(49): 495705, 2012 Dec 14.
Article
in En
| MEDLINE
| ID: mdl-23149480
A new resonance-tracking (RT) method using fast frequency sweeping excitation was developed for quantitative scanning probe microscopy (SPM) imaging. This method allows quantitative imaging of elastic properties and ferroelectrical domains with nanoscale resolution at high data acquisition rates. It consists of a commercial AFM system combined with a high-frequency lock-in amplifier, a programmed function generator and a fast data acquisition card. The resonance-tracking method was applied to the atomic force acoustic microscopy (AFAM) and to the piezoresponse force microscopy (PFM) modes. Plots of amplitude versus time and phase versus time for resonant spectra working with different sweeping frequencies were obtained to evaluate the response speed of the lock-in amplifier. It was proved that this resonance-tracking method allows suitable spectral acquisition at a rate of about 5 ms/pixel, which is useful for SPM imaging in a practical scanning time. In order to demonstrate the system performance, images of RT-AFAM for TiN films and RT-PFM for GeTe are shown.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Image Enhancement
/
Microscopy, Atomic Force
/
Microscopy, Scanning Probe
/
Microscopy, Acoustic
/
Micro-Electrical-Mechanical Systems
Language:
En
Journal:
Nanotechnology
Year:
2012
Document type:
Article
Affiliation country:
Mexico
Country of publication:
United kingdom