INTRODUCTION: The complications of the open-wedge osteotomy technique (slope, hypo- and hypercorrection and fractures) are related to difficulties in the control of the open-wedge angle during surgery. MATERIALS AND METHODS: In this cadaveric study, we evaluated the safety and precision of a novel system, the Realignment High Control System (RHC), in the correction of knee mechanical axis and slope. The RHC has a fixation plate coupled to a dynamic device that opens the osteotomy continuously, allowing plate fixation before osteotomy wedge opening. RESULTS: All procedures were easily performed, with no fractures. The openings equaled the indicated by the navigation system for 7.5° and 10°, and for 5°, there was a statistically, although not clinically significant, difference of 0.6°. The slope shown by the RHC setting and in the navigation system was significantly different for the 10° setting only, with a mean difference of 0.563°. CONCLUSIONS: RHC facilitates the surgical technique of high tibial osteotomy, with gradual wedge opening, precise correction of the mechanical axis, and appropriate control of the tibial slope, even with larger openings.