Accuracy of sizers for aortic valve prostheses.
Bonchek, L I; Burlingame, M W; Vazales, B E.
Affiliation
  • Bonchek LI; Division of Cardiothoracic Surgery, Lancaster General Hospital, Pa.
J Thorac Cardiovasc Surg ; 94(4): 632-4, 1987 Oct.
Article in En | MEDLINE | ID: mdl-3657263
Accurate sizing of the aortic anulus is a prerequisite of prosthetic valve implantation, because it is usually desirable to insert the largest possible prosthesis. Although it is generally assumed that commercial valve sizers are accurately labeled, our measurements reveal that the sizers for the St. Jude Medical, Björk-Shiley, and Ionescu-Shiley prostheses are larger than their nominal sizes. These findings are relevant to the selection of prosthetic valves.