TEG® 6s coagulation testing with a novel heparin neutralization cartridge: Technical validation and determination of normal reference ranges.
Hartmann, Jan; Dias, Joao; Shilo, Alexandra; Bynagari, Yamini; Garrett, Brandon; Jeske, Walter; Manukyan, Zorayr; Mkhitaryan, Karen; Adelmann, Dieter; Subramaniam, Kathirvel; Sakai, Tetsuro.
Affiliation
  • Hartmann J; Global Medical Office, Haemonetics Corporation, Boston, MA, US.
  • Dias J; Global Medical Office, Haemonetics Corporation, Boston, MA, US.
  • Shilo A; Global Medical Office, Haemonetics Corporation, Boston, MA, US.
  • Bynagari Y; Machaon Diagnostics, Inc, Berkeley, CA, US.
  • Garrett B; Machaon Diagnostics, Inc, Berkeley, CA, US.
  • Jeske W; Cardiovascular Research Institute, Loyola University Chicago, Maywood, IL, US.
  • Manukyan Z; ClinStatDevice, Lexington, MA, US.
  • Mkhitaryan K; ClinStatDevice, Lexington, MA, US.
  • Adelmann D; Department of Anesthesia and Perioperative Care, University of California San Francisco, San Francisco, CA, US.
  • Subramaniam K; Department of Anesthesiology and Perioperative Medicine, University of Pittsburgh Medical Center, Pittsburgh, PA, US.
  • Sakai T; Department of Anesthesiology and Perioperative Medicine, University of Pittsburgh Medical Center, Pittsburgh, PA, US.
Am J Clin Pathol ; 2024 Jul 27.
Article in En | MEDLINE | ID: mdl-39073983
ABSTRACT

OBJECTIVES:

We sought to establish normal reference ranges (NRRs) for a novel TEG 6s cartridge (TEG 6s Citrated K, KH, RTH, FFH [Global Hemostasis]) (Haemonetics Corporation, Boston, MA, US).

METHODS:

Healthy volunteers (≥18 years of age) included in this single-arm study provided single samples of whole blood. Primary end points included TEG parameters in the citrated kaolin (CK), CK with heparinase (CKH), RapidTEG with heparinase (CRTH), and functional fibrinogen with heparinase (CFFH) assays.

RESULTS:

Evaluable data were contributed by 164 volunteers (48.8% female; 62% White/Caucasian). The following NRRs were established CK maximum amplitude (MA), 51.0 to 67.6 mm; CKH-MA, 51.8 to 67.9 mm; CRTH-MA, 53.0 to 68.9 mm; CFFH-MA, 15.3 to 34.4 mm; CK reaction time, 5.0 to 9.1 minutes; CKH reaction time, 4.9 to 9.4 minutes; CKH lysis 30 minutes after MA, 0% to 3.2%. Duplicate measurements demonstrated high reproducibility. CFFH-MA correlated with Clauss fibrinogen concentration (Pearson correlation coefficient, 0.74). Laboratory-based studies demonstrated maintenance of the relationship between CFFH-MA and fibrinogen up to 1344 mg/dL (hyperfibrinogenemic samples) and acceptability of heparin neutralization up to concentrations of low molecular weight and unfractionated heparin of 1.3 IU/mL and 5 IU/mL, respectively.

CONCLUSIONS:

This study established NRRs for the Global Hemostasis cartridge and serves as a proof of concept for the validity of results obtained using this cartridge.
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: Am J Clin Pathol Year: 2024 Document type: Article Country of publication: United kingdom