Design and characterization of the time-resolved opacity spectrometer (OpSpecTR) for the NIF iron opacity campaign.
Rev Sci Instrum
; 95(8)2024 Aug 01.
Article
in En
| MEDLINE
| ID: mdl-39101787
ABSTRACT
A new time-resolved opacity spectrometer (OpSpecTR) is currently under development for the National Ignition Facility (NIF) opacity campaign. The spectrometer utilizes Icarus version 2 (IV2) hybridized complementary metal-oxide-semiconductor sensors to collect gated data at the time of the opacity transmission signal, unlocking the ability to collect higher-temperature measurements on NIF. Experimental conditions to achieve higher temperatures are feasible; however, backgrounds will dominate the data collected by the current time-integrating opacity spectrometer. The shortest available OpSpecTR integration time of â¼2 ns is predicted to reduce self-emission and other late-time backgrounds by up to 80%. Initially, three Icarus sensors will be used to collect data in the self-emission, backlighter, and absorption regions of the transmission spectrum, with plans to upgrade to five Daedalus sensors in future implementations with integration times of â¼1.3 ns. We present the details of the diagnostic design along with recent characterization results of the IV2 sensors.