Your browser doesn't support javascript.
loading
Reliability of a pulse oximeter in the detection of hyperoxemia.
Poets, C F; Wilken, M; Seidenberg, J; Southall, D P; von der Hardt, H.
Affiliation
  • Poets CF; Department of Paediatrics, Medizinische Hochschule, Hannover, Germany.
J Pediatr ; 122(1): 87-90, 1993 Jan.
Article in En | MEDLINE | ID: mdl-8419620
To evaluate the reliability of the Nellcor N200 pulse oximeter in the detection of hyperoxemia (oxygen tension > 80 mm Hg), we obtained 213 simultaneous recordings and measurements of transcutaneous oxygen saturation (SO2) and arterial oxygen tension (PaO2) in 50 patients. During 95% of measurements with PaO2 > 80 mm Hg, SO2 was > or = 96%; SO2 was also > or = 96% in 20% of measurements with PaO2 < or = 80 mm Hg (false-positive results for hyperoxemia). With the upper alarm limit set at 95%, the pulse oximeter identified 95% of hyperoxemic instances while allowing PaO2 to be kept to > 60 mm Hg.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Oxygen / Oximetry Type of study: Diagnostic_studies / Evaluation_studies Limits: Child, preschool / Female / Humans / Infant / Male / Newborn Language: En Journal: J Pediatr Year: 1993 Document type: Article Affiliation country: Germany Country of publication: United States
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Oxygen / Oximetry Type of study: Diagnostic_studies / Evaluation_studies Limits: Child, preschool / Female / Humans / Infant / Male / Newborn Language: En Journal: J Pediatr Year: 1993 Document type: Article Affiliation country: Germany Country of publication: United States