Your browser doesn't support javascript.
loading
Two Evaluation Budgets for the Measurement Uncertainty of Glucose in Clinical Chemistry / 대한진단검사의학회지
Article in English | WPRIM (Western Pacific) | ID: wpr-131135
Responsible library: WPRO
ABSTRACT

BACKGROUND:

Measurement uncertainty characterizes the dispersion of the quantity values attributed to a measurand. Although this concept was introduced to medical laboratories some years ago, not all medical researchers are familiar with it. Therefore, the evaluation and expression of measurement uncertainty must be highlighted using a practical example.

METHODS:

In accordance with the procedure for evaluating and expressing uncertainty, provided by the Joint Committee for Guides in Metrology (JCGM), we used plasma glucose (Glu) as an example and defined it as the measurand. We then analyzed the main sources of uncertainty, evaluated each component of uncertainty, and calculated the combined uncertainty and expanded uncertainty with 2 budgets for single measurements and continuous monitoring, respectively.

RESULTS:

During the measurement of Glu, the main sources of uncertainty included imprecision, within-subject biological variance (BVw), calibrator uncertainty, and systematic bias. We evaluated the uncertainty of each component to be 1.26%, 1.91%, 5.70%, 0.42%, and -2.87% for within-run imprecision, between-day imprecision, BVw, calibrator uncertainty, and systematic bias, respectively. For a single specimen, the expanded uncertainty was 7.38% or 6.1+/-0.45 mmol/L (kappa=2); in continuous monitoring of Glu, the expanded uncertainty was 13.58% or 6.1+/-0.83 mmol/L (kappa=2).

CONCLUSIONS:

We have demonstrated the overall procedure for evaluating and reporting uncertainty with 2 different budgets. The uncertainty is not only related to the medical laboratory in which the measurement is undertaken, but is also associated with the calibrator uncertainty and the biological variation of the subject. Therefore, it is helpful in explaining the accuracy of test results.
Subject(s)

Full text: Available Health context: Sustainable Health Agenda for the Americas Health problem: Goal 4: Health financing Database: WPRIM (Western Pacific) Main subject: Quality Control / Blood Chemical Analysis / Models, Statistical / Clinical Chemistry Tests / Uncertainty / Glucose Type of study: Health economic evaluation / Risk factors Limits: Humans Language: English Journal: The Korean Journal of Laboratory Medicine Year: 2011 Document type: Article
Full text: Available Health context: Sustainable Health Agenda for the Americas Health problem: Goal 4: Health financing Database: WPRIM (Western Pacific) Main subject: Quality Control / Blood Chemical Analysis / Models, Statistical / Clinical Chemistry Tests / Uncertainty / Glucose Type of study: Health economic evaluation / Risk factors Limits: Humans Language: English Journal: The Korean Journal of Laboratory Medicine Year: 2011 Document type: Article
...