PURPOSE:

On-line setup procedure was performed before treatment for lung cancer patient for stereotactic radiation surgery (SRS) using recently introduced Cone Beam CT. MATERIALS AND

METHODS:

Cone Beam CT was performed for 10 patients who did SRS during 18 July and 1 September, 2006 using On Board Imager (OPB) system made by Varian, USA. The treatment position of patient was corrected comparing Images obtaining from CBCT and used in treatment planning.

RESULTS:

Comparing the CBCT images and CT images used in treatment planning, the movement of the couch is 2.3+/-2.3 mm, 1.8+/-2.8 mm, 0.3+/-3.5 mm for vertical, lateral, and longitudinal direction. It took about 1 hr for conventional treatment procedure in image acquisition of CT before treatment and image registration. But it took about 4~7 minute in on-line setup using CBCT (1~2 min for image acquisition using CBCT, 2~3 min for CT reconstruction of 2.5 mm slice thickness, 1~2 min for on-line setup using image registration). CONCLUTION The accurate treatment could be performed after tumor localization for SRS using CBCT images. And the consumed time for tumor localization was reduced significantly.
Subject(s)

Full text: Available Database: WPRIM (Western Pacific) Main subject: Cone-Beam Computed Tomography / Radiotherapy, Image-Guided / Lung / Lung Neoplasms Limits: Humans Language: Korean Journal: Journal of Lung Cancer Year: 2006 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Main subject: Cone-Beam Computed Tomography / Radiotherapy, Image-Guided / Lung / Lung Neoplasms Limits: Humans Language: Korean Journal: Journal of Lung Cancer Year: 2006 Document type: Article
...