Evaluation of Kanto Cica Liquid Reagents for Chemistry Analyzer / 임상검사와정도관리
Article in Korean | WPRIM (Western Pacific) | ID: wpr-210981
Responsible library: WPRO
ABSTRACT

BACKGROUND:

Many reagents have been developed along with advances in chemistry auto analyzer. Deciding on an appropriate reagent is required for an accurate test, diagnosis and efficient laboratory management. We evaluated Cica Liquid reagent produced by Kanto chemical corporation (Tokyo, Japan) for checking reagent ability.

METHODS:

Twelve chemistry reagents (AST, ALT, ALP, glucose, BUN, creatinine, total bilirubin, direct bilirubin, cholesterol, triglyceride, iron, magnesium) were tested on precision, linearity, interference, and correlation. We have evaluated using Hitachi 7600 (Hitachi High Technologies co., Japan) chemistry auto analyzer in accordance with the CLSI guidelines EP5-A, EP6-A, EP7-A, EP9-A2. EP_Suite (Marchem Associates Inc., USA) and SPSS ver. 11.0 (SPSS Inc., USA) were used for statistics.

RESULTS:

Precision results were satisfactory to CLIA '88 in all of the analytes except for ALP and magnesium. The linearity was satisfactory in measurement ranges as all analytes showed linearity in polynomial regression result or relative nonlinearity of less than 2.5%. Coefficients of correlation were above 0.985 in all analytes except for direct bilirubin and magnesium. When interference test results were compared with criteria of CLIA '88, low level of AST was positively interfered by hemoglobin and magnesium was negatively interfered by bilirubin.

CONCLUSIONS:

In conclusion, the Kanto Cica Liquid reagents are valuable in clinical laboratory, since they showed good precision, linearity, and correlation with other reagents.
Subject(s)

Full text: Available Database: WPRIM (Western Pacific) Main subject: Triglycerides / Bilirubin / Chemistry / Cholesterol / Creatinine / Diagnosis / Glucose / Indicators and Reagents / Iron / Magnesium Type of study: Diagnostic study / Practice guideline Language: Korean Journal: Journal of Laboratory Medicine and Quality Assurance Year: 2007 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Main subject: Triglycerides / Bilirubin / Chemistry / Cholesterol / Creatinine / Diagnosis / Glucose / Indicators and Reagents / Iron / Magnesium Type of study: Diagnostic study / Practice guideline Language: Korean Journal: Journal of Laboratory Medicine and Quality Assurance Year: 2007 Document type: Article
...