Your browser doesn't support javascript.
loading
Improvement of Digital Capsule Endoscopy System and Image Interpolation / 中国医疗器械杂志
Article in Chinese | WPRIM (Western Pacific) | ID: wpr-265581
Responsible library: WPRO
ABSTRACT
Traditional capsule image collects and transmits analog image, with weak anti-interference ability, low frame rate, low resolution. This paper presents a new digital image capsule, which collects and transmits digital image, with frame rate up to 30 frames/sec and pixels resolution of 400 x 400. The image is compressed in the capsule, and is transmitted to the outside of the capsule for decompression and interpolation. A new type of interpolation algorithm is proposed, which is based on the relationship between the image planes, to obtain higher quality colour images. capsule endoscopy, digital image, SCCB protocol, image interpolation
Subject(s)
Full text: Available Database: WPRIM (Western Pacific) Main subject: Algorithms / Image Processing, Computer-Assisted / Equipment Design / Capsule Endoscopy Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2016 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Main subject: Algorithms / Image Processing, Computer-Assisted / Equipment Design / Capsule Endoscopy Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2016 Document type: Article
...