Your browser doesn't support javascript.
loading
Study on Accurately Controlling Discharge Energy Method Used in External Defibrillator / 中国医疗器械杂志
Article in Chinese | WPRIM (Western Pacific) | ID: wpr-265583
Responsible library: WPRO
ABSTRACT
This paper introduces a new method which controls discharge energy accurately. It is achieved by calculating target voltage based on transthoracic impedance and accurately controlling charging voltage and discharge pulse width. A new defibrillator is designed and programmed using this method. The test results show that this method is valid and applicable to all kinds of external defibrillators.
Subject(s)
Full text: Available Database: WPRIM (Western Pacific) Main subject: Reference Standards / Defibrillators / Equipment Design Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2016 Document type: Article