Determination of Fe,Cu,Zn,Mn in Pollen by Flame Atomic Absorption Spectrophotometry / 环境与健康杂志
Article in Chinese | WPRIM (Western Pacific) | ID: wpr-547650
Responsible library: WPRO
ABSTRACT
Objective To establish a method for the determination of Fe, Cu, Zn, Mn in the pollen by flame atomic absorption spectrometry(FAAS).Methods Wet digestion-flame atomic absorption spectrophotometry was used to determine the content of Fe, Cu, Zn, Mn in the pollen.Results In the selected experimental conditions, the recovery rates of these elements were 98.5%-103.0% and the coefficient of variation was 1.13%-2.36%.The detection limits were 0.40 ?g/L for Fe, 0.08 ?g/L for Cu, 0.20 ?g/L for Zn, 0.14 ?g/L for Mn.The linear range was 0.00-2.00 mg/L.Conclusion FAAS is simple, fast, accurate for the determination of Fe, Cu, Zn, Mn in the pollen.

Full text: Available Database: WPRIM (Western Pacific) Language: Chinese Journal: Journal of Environment and Health Year: 1989 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Language: Chinese Journal: Journal of Environment and Health Year: 1989 Document type: Article
...