Your browser doesn't support javascript.
loading
Measurement of output dose of medical linear accelerator with IAEA regulation / 医疗卫生装备
Article in Chinese | WPRIM (Western Pacific) | ID: wpr-586672
Responsible library: WPRO
ABSTRACT
Objective To probe an accurate,rapid and effective method for the calculation and measurement of output dose of the accelerator with IAEA regulation.Methods The parameter value of measurement was determined according to IAEA regulations,and absorbed dose calculation forma was briefed to the form that the reading of instrument multiplies cumulative correction factor.The output dose was measured in water phantom on sandard condition,and then regrafted into plexiglass phantom,which made the measurement more convenience.Results Staistics of measurement had been done to PRIMUS accelerator in 66 weeks output dose.Over 85% of the measurement had the error less then 1%,and others were less then 2%.Conclusion This measurement can be used for the measuremen of output dose in precise radiotherapy of linear accelerator.

Full text: Available Database: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Medical Equipment Journal Year: 1993 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Medical Equipment Journal Year: 1993 Document type: Article
...