Performance Tests and Quality Control of Siemens ECAT EXACT HR+ PET / 医疗卫生装备
Chinese Medical Equipment Journal
; (6)2004.
Article
in Chinese
| WPRIM (Western Pacific)
| ID: wpr-592091
Responsible library:
WPRO
ABSTRACT
Objective To evaluate the performance parameters of the PET instrument, which has been used for years in our hospital, assuring the stability of image quality and performance parameters. Methods NEMA NU 2-1994 Standards was used as reference for performance measurement of PET, which were published by National Electrical Manufacturers Association(NEMA). The NEMA standard cylinder model, 18F radiation source, and 18F and 68Ge columniform radiation source were used. Results Projects for performance measurement and daily quality control were drawn up. Conclusion This project provides a reference and methods for the performance measurement of PET instrument and daily quality control. It better ensures the stability of image quality and performance parameters of the PET instrument.
Full text:
Available
Database:
WPRIM (Western Pacific)
Language:
Chinese
Journal:
Chinese Medical Equipment Journal
Year:
2004
Document type:
Article