Development of standard defibrillation energy source / 医疗卫生装备
Chinese Medical Equipment Journal
; (6): 10-13,33, 2015.
Article
in Chinese
| WPRIM (Western Pacific)
| ID: wpr-602921
Responsible library:
WPRO
ABSTRACT
To develop a standard defibrillation energy source which can output monophasic and biphasic de-fibrillation impulses and can display standard energy values. The storage capacitor was charged by single-end flyback transformator and discharged through the H bridge composed of insulated gate bipolar transistor (IGBT). Impulse voltage and current were collected by divider while discharging, and the energy value was calculated by STM32. The en-ergy value and waveform were displayed through the control module. The energy source could output both monophsic and biphasic waveforms, and the accuracy of displayed value was higher than ±2% or ±1 J. The standard defibrillation energy source can be used as standard device for defibrillator analyzer, and the metrological trace-ability system of defibrillation energy may come to be completed.
Full text:
Available
Database:
WPRIM (Western Pacific)
Language:
Chinese
Journal:
Chinese Medical Equipment Journal
Year:
2015
Document type:
Article