Your browser doesn't support javascript.
loading
Intraoperative Measurement of Femorotibial Contact Pressure in Total Knee Arthroplasty / 대한정형외과학회잡지
Article in Korean | WPRIM (Western Pacific) | ID: wpr-649967
Responsible library: WPRO
ABSTRACT

PURPOSE:

A new method has been devised to determine contact pressures in prosthetic components. A real time analysis of the contact pressure distribution could help fine tuned implant positioning and ligament balancing during Total Knee Arthroplasty. MATERIALS AND

METHODS:

Posterior cruciate reserving type procedures were done with externally rotated femoral cut. Femorotibial contact pressure was measured in four compartments in both flexion and extension position. A LOAD CELL located beneath the tibial plate changes the contact pressure into electric signal and DP-41S transformed the electric sign into the pressure unit. Adjunctive fine-tuning operation was done to approximate the equilibrium among the four compartments of the tibial plate.

RESULTS:

Femorotibial contact pressure were distributed between 17.1 N (Newton) to 33.4 N (average 23.6 N). A peak contact pressure was exhibited in the anteromedial compartment in extended knee and in the posterolateral compartment in flexed knee.

CONCLUSION:

A LOAD CELL electronic transducer technique was introduced. This method was tried to get the balanced equilibrium of contact pressure between Femorotibial components during TKA.
Subject(s)

Full text: Available Database: WPRIM (Western Pacific) Main subject: Arthroplasty / Transducers / Knee / Ligaments Language: Korean Journal: The Journal of the Korean Orthopaedic Association Year: 2001 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Main subject: Arthroplasty / Transducers / Knee / Ligaments Language: Korean Journal: The Journal of the Korean Orthopaedic Association Year: 2001 Document type: Article
...