Application of EXCEL in Evaluating the Precision Performance of XE-2100 Automatic Blood Cell Analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine
; (12): 3240-3242, 2017.
Article
in Chinese
| WPRIM (Western Pacific)
| ID: wpr-664082
Responsible library:
WPRO
ABSTRACT
Objective To evaluation the precision performance of XE-2100 Automatic Blood Cell Analyzer using EXCEL Mi-crosoft with reference to CLSI EP15-A3 .Methods According to CLSI EP15-A3 ,high level and low level (BIO-RAD company) of quality control materials were measured using XE-2100 Automatic Blood Cell Analyzer ,and analyzed by one-factor analysis of vari-ance of EXCEL 2003 and equation editor .The within-run coefficient of variation and within-laboratory coefficient of variation were calculated and compared with national industry standards and manufacturer statement .Results For high and low level of quality control materials ,the within-run coefficient of variation and within-laboratory coefficient of variation of white blood cell ,red blood cell ,hemoglobin and platelet were less than national industry standards and manufacturer statement .Conclusion The precision ver-ification scheme of CLSI EP15-A3 with EXCEL Microsoft has a relatively good maneuverability and statistical performance ,which could be used in verification of the precision performance of quantitative testing items in clinical laboratories .
Full text:
Available
Database:
WPRIM (Western Pacific)
Type of study:
Practice guideline
Language:
Chinese
Journal:
International Journal of Laboratory Medicine
Year:
2017
Document type:
Article