Your browser doesn't support javascript.
loading
Investigation and practice of magnetic field homogeneity evaluation methods for MRI equipment / 中国医学影像技术
Article in Chinese | WPRIM (Western Pacific) | ID: wpr-668735
Responsible library: WPRO
ABSTRACT
Objective To explore the evaluation methods of magnetic field uniformity and the practical operations,in order to provide references for clinical quality control.Methods Based on MRS phantom and 32 cm uniform spherical phantom,the evaluation methods of magnetic field uniformity were used to detect MRI equipment in department of radiology,including spectral peak and the bandwidth difference.Results The full width at half maximum (FWHM) of MRS phantom was 0.08 ppm,the inhomogeneity values on coronal,sagittal and axial detected by the bandwidth difference method was 0.15 ppm,0.08 ppm and 0.18 ppm,respectively.Conclusion The magnetic field homogeneity of the above mentioned MRI equipment is favorable,which accords with magnetic field uniformity detection standard.

Full text: Available Database: WPRIM (Western Pacific) Type of study: Evaluation study Language: Chinese Journal: Chinese Journal of Medical Imaging Technology Year: 2017 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Type of study: Evaluation study Language: Chinese Journal: Chinese Journal of Medical Imaging Technology Year: 2017 Document type: Article
...