Clinical Utility and Cross-Reactivity of Insulin and C-Peptide Assays by the Lumipulse G1200 System
Annals of Laboratory Medicine
; : 530-537, 2018.
Article
in English
| WPRIM (Western Pacific)
| ID: wpr-718330
Responsible library:
WPRO
ABSTRACT
BACKGROUND:
Measurement of insulin and C-peptide concentrations is important for deciding whether insulin treatment is required in diabetic patients. We aimed to investigate the analytical performance of insulin and C-peptide assays using the Lumipulse G1200 system (Fujirebio Inc., Tokyo, Japan).METHODS:
We examined the precision, linearity, and cross-reactivity of insulin and C-peptide using five insulin analogues and purified proinsulin. A method comparison was conducted between the Lumipulse G1200 and Roche E170 (Roche Diagnostics, Mannheim, Germany) systems in 200 diabetic patients on insulin treatment. Reference intervals for insulin and C-peptide concentrations were determined in 279 healthy individuals.RESULTS:
For insulin and C-peptide assays, within-laboratory precision (% CV) was 3.78–4.14 and 2.89–3.35%, respectively. The linearity of the insulin assay in the range of 0–2,778 pmol/L was R2=0.9997, and that of the C-peptide assay in the range of 0–10 nmol/L was R2=0.9996. The correlation coefficient (r) between the Roche E170 and Lumipulse G1200 results was 0.943 (P < 0.001) for insulin and 0.996 (P < 0.001) for C-peptide. The mean differences in insulin and C-peptide between Lumipulse G1200 and the Roche E170 were 19.4 pmol/L and 0.2 nmol/L, respectively. None of the insulin analogues or proinsulin showed significant cross-reactivity with the Lumipulse G1200. Reference intervals of insulin and C-peptide were 7.64–70.14 pmol/L and 0.17–0.85 nmol/L, respectively.CONCLUSIONS:
Insulin and C-peptide tests on the Lumipulse G1200 show adequate analytical performance and are expected to be acceptable for use in clinical areas.
Full text:
Available
Database:
WPRIM (Western Pacific)
Main subject:
Proinsulin
/
C-Peptide
/