Your browser doesn't support javascript.
loading
Repeated endotracheal tube cuff tears during nasotracheal intubation due to nasal cavity orthodontic micro-implant - A case report -
Article in English | WPRIM (Western Pacific) | ID: wpr-913378
Responsible library: WPRO
ABSTRACT
Background@#Nasotracheal intubation is generally performed for intraoral surgery.Case A 34-year-old female patient who underwent orthognathic surgery exhibited repeated endotracheal tube cuff tears during nasotracheal intubation. After intubation, leaks developed, and torn endotracheal cuff was observed in the removed endotracheal tube. Subsequently, re-intubation through the same nasal cavity was performed immediately, but leakage from the torn endotracheal tube cuff was re-observed. A leakage test of the extubated tube revealed air bubbles and leaks near the tube cuff due to the tear. Nasotracheal intubation was performed through the other nasal cavity, and there were no leakage findings or abnormalities. During the course of the surgery, the surgeon noticed that the orthodontic micro-implant deposited in the mid-tube cavity was exposed to the nasal cavity. @*Conclusions@#We aimed to emphasize caution and discuss the possibility that orthodontic micro-implants that are not confirmed during preoperative evaluation may cause repeated endotracheal tube cuff tears.
Full text: Available Database: WPRIM (Western Pacific) Language: English Journal: Anesthesia and Pain Medicine Year: 2021 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Language: English Journal: Anesthesia and Pain Medicine Year: 2021 Document type: Article
...