Study on Influencing Factors of the Tip Softness of Epidural Anesthesia Catheter / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
; (6): 483-486, 2021.
Article
in Zh
| WPRIM
| ID: wpr-922043
Responsible library:
WPRO
ABSTRACT
This article aims to study the factors affecting the flexibility of the tip of an epidural anesthesia catheter. The flexibility of the tip of the epidural anesthesia catheter was tested with a softness tester from four aspects:raw materials, tip structure, tip processing technology, and the outer diameter of the catheter. Highly flexible and malleable polymer material with a smooth tip, the tip softening process and the proper outer diameter can effectively improve the tip flexibility of the epidural anesthesia catheter.
Key words
Full text:
1
Database:
WPRIM
Main subject:
Catheterization
/
Epidural Space
/
Catheters
/
Anesthesia, Epidural
Language: