Your browser doesn't support javascript.
loading
Simulation Study on the Influence of Sampling Delay on the Accuracy of Energy Metabolism Measurement / 中国医疗器械杂志
Article in Chinese | WPRIM (Western Pacific) | ID: wpr-939752
Responsible library: WPRO
ABSTRACT
Indirect energy metabolism measurement is the gold standard for providing nutritional support for critical illness. The accuracy of the measurement data directly affects the outcome of the disease. In order to study the influence of sampling delay on the accuracy of energy metabolism measurement under mechanical ventilation, the Matlab/Simulink platform and respiratory electrical model were used for simulation and quantitative analysis. The results show that the error of indirect energy metabolism measurement increases with the increase of sampling delay, the error of sampling delay in mechanical ventilation mode is larger than that of spontaneous breathing, and the error of sampling delay in PCV mode of mechanical ventilation is larger than that in VCV mode. Therefore, there should be different sampling delay compensation strategies under severe mechanical ventilation and its different control modes.
Subject(s)

Full text: Available Database: WPRIM (Western Pacific) Main subject: Respiration, Artificial / Computer Simulation / Critical Illness / Energy Metabolism Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2022 Document type: Article
Full text: Available Database: WPRIM (Western Pacific) Main subject: Respiration, Artificial / Computer Simulation / Critical Illness / Energy Metabolism Limits: Humans Language: Chinese Journal: Chinese Journal of Medical Instrumentation Year: 2022 Document type: Article
...