Experimental Study of MAR Algorithm in Metal Artifact Removal of CT Simulator / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
; (6): 24-27, 2020.
Article
in Chinese
| WPRIM (Western Pacific)
| ID: wpr-942690
Responsible library:
WPRO
ABSTRACT
OBJECTIVE@#To explore the application value of MAR algorithm in metal artifact removal of CT simulator.@*METHODS@#CT phantom with titanium plate was scanned using conventional algorithms and MAR algorithms, respectively. Artifact index(AI), contrast-to-noise ratio(CNR) and AI values at different slices were used to analyze the artifact images.@*RESULTS@#In artifact index, MAR algorithm (10.28±2.60) is significantly lower than conventional algorithm (20.65±5.04); In contrast-to-noise ratio index, MAR algorithm (7.81±1.12) is better than conventional algorithm (5.61±1.36). The above indicators were statistically significant in both algorithms (P<0.01). In the slices affected by metal artifacts, the artifact index decreased by 21.72%~88.40% after the MAR algorithm.@*CONCLUSIONS@#MAR algorithm can significantly reduce the metal artifacts and improve the clinical value of CT data.
Full text:
Available
Database:
WPRIM (Western Pacific)
Main subject:
Titanium
/
Algorithms
/
Tomography, X-Ray Computed
/
Artifacts
/
Phantoms, Imaging
/
Metals
Type of study:
Prognostic study
Language:
Chinese
Journal:
Chinese Journal of Medical Instrumentation
Year:
2020
Document type:
Article