Extremely high dose neutron dosimetry using CR-39 and atomic force microscopy.
Radiat Prot Dosimetry
; 120(1-4): 470-4, 2006.
Article
en En
| MEDLINE
| ID: mdl-16597695
Atomic force microscopy (AFM) has been applied to the analysis of CR-39 nuclear track detectors for high dose neutron dosimetry. As a feasible study to extract the neutron dose, we have employed a (239)Pu-Be neutron source with the traditional track density measurement of recoil proton etch pits from a high density polyethylene (CH(2)) radiator. After very short etching ( approximately 1 microm), etch pit densities were measured as a function of neutron fluence (neutron dose) up to 1.4 x 10(10) cm(-2) (6.6 Sv). Neutron sensitivity was also measured to be 6.6 x 10(-4). Maximum measurable neutron dose was estimated to be approximately 200 Sv by measuring the fraction of the total image area occupied by the etch pits.