Characterisation of kilo electron volt neutron fluence standard with the 45Sc(p,n)45Ti reaction at NMIJ.
Radiat Prot Dosimetry
; 126(1-4): 155-8, 2007.
Article
en En
| MEDLINE
| ID: mdl-17513857
We are developing a national standard of a monoenergetic kilo electron volt neutron field with the (45)Sc(p,n)(45)Ti resonance reaction. A wide resonance yields 27.4 keV neutrons at 0 degrees with respect to the proton beam. The proton energy was precisely determined in the measurement of the relative neutron yield as a function of the proton energy from the threshold energy to 2.942 MeV. Absolute measurement of the monoenergetic neutron fluence was performed using a (3)He proportional counter. Relative measurement was also carried out using a Bonner sphere calibrated at our 144 keV standard neutron field. Calibration factors were obtained between the count of a neutron monitor and the neutron fluence. A silicon-surface barrier detector with a (6)LiF foil converter was also being developed for the neutron fluence measurement. Successful results were obtained in the tests in the 144 keV standard neutron field.
Buscar en Google
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Radiometría
/
Escandio
/
Titanio
/
Neutrones
País/Región como asunto:
Asia
Idioma:
En
Revista:
Radiat Prot Dosimetry
Año:
2007
Tipo del documento:
Article
País de afiliación:
Japón
Pais de publicación:
Reino Unido