Wide-field turbulence imaging with beam emission spectroscopy.
McKee, G R; Fonck, R J; Shafer, M W; Uzun-Kaymak, I U; Yan, Z.
Afiliación
  • McKee GR; University of Wisconsin-Madison, Madison, Wisconsin 53706, USA. grmckee@wisc.edu
Rev Sci Instrum ; 81(10): 10D741, 2010 Oct.
Article en En | MEDLINE | ID: mdl-21033932
Imaging of the size, shape, time-averaged, and time-resolved dynamics of long-wavelength density turbulence structures is accomplished with an expanded, high-sensitivity, wide-field beam emission spectroscopy (BES) diagnostic on DIII-D. A 64-channel BES system is configured with an 8×8 grid of discrete channels that image an approximately 7×9 cm region at the outboard midplane. The grid covers multiple correlation lengths and each channel shape matches the measured radial-poloidal correlation length asymmetry of turbulent eddies. The wide field 8×8 imaging capability allows for sampling of essentially the full two-dimensional spatial correlation function for typical plasma conditions. The sampled area can be radially scanned over 0.4

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Rev Sci Instrum Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos