Development and validation of a dried blood spot LC-MS/MS assay to quantify ranitidine in paediatric samples.
J Pharm Biomed Anal
; 56(5): 1057-63, 2011 Dec 15.
Article
en En
| MEDLINE
| ID: mdl-21890296
A novel approach has been developed to determine ranitidine in paediatric samples using dried blood spots (DBS) on Guthrie cards (Whatman 903). A selective and sensitive HPLC-MS/MS assay has been developed and validated using small volumes of blood (30 µl). A 6 mm disc was punched from each DBS and extracted with methanolic solution of the internal standard (IS) nizatidine. This was further subjected to solid phase extraction (SPE), followed by reversed phase HPLC separation, using a XBridge™ C18 column and mobile phase 10 mM ammonium acetate/methanol (98:2 v/v) with a flow rate of 0.3 mL/min. This was combined with multiple reaction monitoring (MRM) mass detection using electrospray ionisation (ESI). The calibration curve for ranitidine was found linear over the range 10-500 ng/mL (r=0.996). The limit of quantification (LOQ) of the method was validated at 10 ng/mL. Accuracy and precision values for within and between days were <20% at the LOQ and <15% at all other concentrations. The validated DBS method was successfully applied to a clinical study employing 81 samples from 36 paediatric patients.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Ranitidina
/
Cromatografía Líquida de Alta Presión
/
Espectrometría de Masa por Ionización de Electrospray
/
Espectrometría de Masas en Tándem
/
Antagonistas de los Receptores H2 de la Histamina
Límite:
Child
/
Humans
Idioma:
En
Revista:
J Pharm Biomed Anal
Año:
2011
Tipo del documento:
Article
País de afiliación:
Reino Unido
Pais de publicación:
Reino Unido