(31)P CSI of the human brain in healthy subjects and tumor patients at 9.4 T with a three-layered multi-nuclear coil: initial results.
Mirkes, Christian; Shajan, Gunamony; Chadzynski, Grzegorz; Buckenmaier, Kai; Bender, Benjamin; Scheffler, Klaus.
Afiliación
  • Mirkes C; Department for Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany. christian.mirkes@tuebingen.mpg.de.
  • Shajan G; High-Field MR Center, Max Planck Institute for Biological Cybernetics, Spemannstr. 41, 72076, Tübingen, Germany. christian.mirkes@tuebingen.mpg.de.
  • Chadzynski G; High-Field MR Center, Max Planck Institute for Biological Cybernetics, Spemannstr. 41, 72076, Tübingen, Germany.
  • Buckenmaier K; Department for Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany.
  • Bender B; High-Field MR Center, Max Planck Institute for Biological Cybernetics, Spemannstr. 41, 72076, Tübingen, Germany.
  • Scheffler K; High-Field MR Center, Max Planck Institute for Biological Cybernetics, Spemannstr. 41, 72076, Tübingen, Germany.
MAGMA ; 29(3): 579-89, 2016 Jun.
Article en En | MEDLINE | ID: mdl-26811174
ABSTRACT

OBJECTIVE:

Investigation of the feasibility and performance of phosphorus ((31)P) magnetic resonance spectroscopic imaging (MRSI) at 9.4 T with a three-layered phosphorus/proton coil in human normal brain tissue and tumor. MATERIALS AND

METHODS:

A multi-channel (31)P coil was designed to enable MRSI of the entire human brain. The performance of the coil was evaluated by means of electromagnetic field simulations and actual measurements. A 3D chemical shift imaging approach with a variable repetition time and flip angle was used to increase the achievable signal-to-noise ratio of the acquired (31)P spectra. The impact of the resulting k-space modulation was investigated by simulations. Three tumor patients and three healthy volunteers were scanned and differences between spectra from healthy and cancerous tissue were evaluated qualitatively.

RESULTS:

The high sensitivity provided by the 27-channel (31)P coil allowed acquiring CSI data in 22 min with a nominal voxel size of 15 × 15 × 15 mm(3). Shimming and anatomical localization could be performed with the integrated four-channel proton dipole array. The amplitudes of the phosphodiesters and phosphoethanolamine appeared reduced in tumorous tissue for all three patients. A neutral or slightly alkaline pH was measured within the brain lesions.

CONCLUSION:

These initial results demonstrate that (31)P 3D CSI is feasible at 9.4 T and could be performed successfully in healthy subjects and tumor patients in under 30 min.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Encéfalo / Neoplasias Encefálicas / Imagen por Resonancia Magnética Tipo de estudio: Prognostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: MAGMA Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2016 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Encéfalo / Neoplasias Encefálicas / Imagen por Resonancia Magnética Tipo de estudio: Prognostic_studies Límite: Adult / Female / Humans / Male Idioma: En Revista: MAGMA Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2016 Tipo del documento: Article País de afiliación: Alemania