Non-invasive continuous arterial pressure and pulse pressure variation measured with Nexfin(®) in patients following major upper abdominal surgery: a comparative study.
Anaesthesia
; 71(7): 788-97, 2016 07.
Article
en En
| MEDLINE
| ID: mdl-27291598
We compared the accuracy and precision of the non-invasive Nexfin(®) device for determining systolic, diastolic, mean arterial pressure and pulse pressure variation, with arterial blood pressure values measured from a radial artery catheter in 19 patients following upper abdominal surgery. Measurements were taken at baseline and following fluid loading. Pooled data results of the arterial blood pressures showed no difference between the two measurement modalities. Bland-Altman analysis of pulse pressure variation showed significant differences between values obtained from the radial artery catheter and Nexfin finger cuff technology (mean (SD) 1.49 (2.09)%, p < 0.001, coefficient of variation 24%, limits of agreement -2.71% to 5.69%). The effect of volume expansion on pulse pressure variation was identical between methods (concordance correlation coefficient 0.848). We consider the Nexfin monitor system to be acceptable for use in patients after major upper abdominal surgery without major cardiovascular compromise or haemodynamic support.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Cuidados Posoperatorios
/
Presión Sanguínea
/
Determinación de la Presión Sanguínea
/
Monitores de Presión Sanguínea
/
Abdomen
Límite:
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Anaesthesia
Año:
2016
Tipo del documento:
Article
País de afiliación:
Países Bajos
Pais de publicación:
Reino Unido