Reduction in Radiation Dose in a Pediatric Cardiac Catheterization Lab Using the Philips AlluraClarity X-ray System.
Sullivan, Patrick M; Harrison, David; Badran, Sarah; Takao, Cheryl M; Ing, Frank F.
Afiliación
  • Sullivan PM; Division of Pediatric Cardiology, Children's Hospital Los Angeles, University of Southern California Keck School of Medicine, Los Angeles, CA, USA. patrick.morris.sullivan@gmail.com.
  • Harrison D; Children's Hospital Los Angeles Heart Center, 4650 Sunset Blvd., #34, Los Angeles, CA, USA. patrick.morris.sullivan@gmail.com.
  • Badran S; Division of General Pediatrics, Children's Hospital Los Angeles, University of Southern California Keck School of Medicine, Los Angeles, CA, USA.
  • Takao CM; Division of Pediatric Cardiology, Children's Hospital Los Angeles, University of Southern California Keck School of Medicine, Los Angeles, CA, USA.
  • Ing FF; Division of Pediatric Cardiology, Children's Hospital Los Angeles, University of Southern California Keck School of Medicine, Los Angeles, CA, USA.
Pediatr Cardiol ; 38(8): 1583-1591, 2017 Dec.
Article en En | MEDLINE | ID: mdl-28770307
The objective of this study was to compare radiation doses and imaging quality using Philips AlluraClarity (Philips Healthcare, Best, The Netherlands) X-ray system and an older generation reference system. AlluraClarity is a new generation fluoroscopy system designed to reduce radiation without compromising image quality, but reports of its use in pediatric patients are limited. Dose area products (DAP, mGy cm2) and DAP/kg were compared in patients catheterized using Allura Xper and AlluraClarity systems over a year of use for each. Randomly selected studies from each system were assessed for image quality. The 430 patients imaged with Clarity were larger than the 332 imaged with Xper (median BSA: 0.74 vs. 0.64 m2, p = 0.06), and median total fluoroscopic times (TFT) were similar (15.8 vs. 16.1 min, p = 0.37). Median DAPs were 8661 mGy cm2 (IQR: 18,300 mGy cm2) and 4523 mGy cm2 (IQR: 11,596 mGy cm2) with Xper and Clarity, respectively (p < 0.001). There was a reduction in median DAP in all procedure categories. After adjustment for BSA, TFT, and procedure type, using Clarity was associated with a 57.5% (95% CI 51.5-62.8%, p < 0.001) reduction in DAP for all procedures. Reductions did not significantly differ by weight (<10 kg, 10-40 kg, ≥ 40 kg). There was an adjusted percent reduction in DAP for each procedure category ranging from 39.0% (95% CI 25.6-50.1%, p < 0.001) for cardiac biopsies with or without coronary angiography to 67.6% (95% CI 61.2-72.8%, p < 0.001) for device occlusions. Mean overall imaging quality scores (4.3 ± 0.8 with Clarity vs. 4.4 ± 0.6 with Xper, p = 0.62) and scores based on specific quality parameters were similar in the two groups. Use of AlluraClarity substantially reduced radiation doses compared to the older generation reference system without compromising imaging quality in a pediatric cardiac catheterization lab.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Fluoroscopía / Cateterismo Cardíaco / Exposición a la Radiación Tipo de estudio: Etiology_studies / Risk_factors_studies Límite: Adolescent / Adult / Child / Child, preschool / Humans / Infant / Newborn País/Región como asunto: Europa Idioma: En Revista: Pediatr Cardiol Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Fluoroscopía / Cateterismo Cardíaco / Exposición a la Radiación Tipo de estudio: Etiology_studies / Risk_factors_studies Límite: Adolescent / Adult / Child / Child, preschool / Humans / Infant / Newborn País/Región como asunto: Europa Idioma: En Revista: Pediatr Cardiol Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Estados Unidos