On the accuracy of optically tracked transducers for image-guided transcranial ultrasound.
Chaplin, V; Phipps, M A; Jonathan, S V; Grissom, W A; Yang, P F; Chen, L M; Caskey, C F.
Afiliación
  • Chaplin V; Department of Radiology and Radiological Sciences, Institute of Imaging Science, Vanderbilt University Medical Center, AA 1105 MCN, 1161 21st Ave. S, Nashville, TN, TN 37232, USA.
  • Phipps MA; Department of Radiology and Radiological Sciences, Institute of Imaging Science, Vanderbilt University Medical Center, AA 1105 MCN, 1161 21st Ave. S, Nashville, TN, TN 37232, USA.
  • Jonathan SV; Biomedical Engineering, Vanderbilt University, Nashville, TN, USA.
  • Grissom WA; Department of Radiology and Radiological Sciences, Institute of Imaging Science, Vanderbilt University Medical Center, AA 1105 MCN, 1161 21st Ave. S, Nashville, TN, TN 37232, USA.
  • Yang PF; Biomedical Engineering, Vanderbilt University, Nashville, TN, USA.
  • Chen LM; Department of Radiology and Radiological Sciences, Institute of Imaging Science, Vanderbilt University Medical Center, AA 1105 MCN, 1161 21st Ave. S, Nashville, TN, TN 37232, USA.
  • Caskey CF; Department of Radiology and Radiological Sciences, Institute of Imaging Science, Vanderbilt University Medical Center, AA 1105 MCN, 1161 21st Ave. S, Nashville, TN, TN 37232, USA.
Int J Comput Assist Radiol Surg ; 14(8): 1317-1327, 2019 Aug.
Article en En | MEDLINE | ID: mdl-31069643
PURPOSE: Transcranial focused ultrasound (FUS) is increasingly being explored to modulate neuronal activity. To target neuromodulation, researchers often localize the FUS beam onto the brain region(s) of interest using spatially tracked tools overlaid on pre-acquired images. Here, we quantify the accuracy of optically tracked image-guided FUS with magnetic resonance imaging (MRI) thermometry, evaluate sources of error and demonstrate feasibility of these procedures to target the macaque somatosensory region. METHODS: We developed an optically tracked FUS system capable of projecting the transducer focus onto a pre-acquired MRI volume. To measure the target registration error (TRE), we aimed the transducer focus at a desired target in a phantom under image guidance, heated the target while imaging with MR thermometry and then calculated the TRE as the difference between the targeted and heated locations. Multiple targets were measured using either an unbiased or bias-corrected calibration. We then targeted the macaque S1 brain region, where displacement induced by the acoustic radiation force was measured using MR acoustic radiation force imaging (MR-ARFI). RESULTS: All calibration methods enabled registration with TRE on the order of 3 mm. Unbiased calibration resulted in an average TRE of 3.26 mm (min-max: 2.80-4.53 mm), which was not significantly changed by prospective bias correction (TRE of 3.05 mm; 2.06-3.81 mm, p = 0.55). Restricting motion between the transducer and target and increasing the distance between tracked markers reduced the TRE to 2.43 mm (min-max: 0.79-3.88 mm). MR-ARFI images showed qualitatively similar shape and extent as projected beam profiles in a living non-human primate. CONCLUSIONS: Our study describes methods for image guidance of FUS neuromodulation and quantifies errors associated with this method in a large animal. The workflow is efficient enough for in vivo use, and we demonstrate transcranial MR-ARFI in vivo in macaques for the first time.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Transductores / Encéfalo / Imagen por Resonancia Magnética / Ultrasonografía / Neuroimagen Tipo de estudio: Diagnostic_studies / Observational_studies / Risk_factors_studies Límite: Animals Idioma: En Revista: Int J Comput Assist Radiol Surg Asunto de la revista: RADIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Transductores / Encéfalo / Imagen por Resonancia Magnética / Ultrasonografía / Neuroimagen Tipo de estudio: Diagnostic_studies / Observational_studies / Risk_factors_studies Límite: Animals Idioma: En Revista: Int J Comput Assist Radiol Surg Asunto de la revista: RADIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Estados Unidos Pais de publicación: Alemania