Your browser doesn't support javascript.
loading
Long-term impact of sagittal malalignment on hardware after posterior fixation of the thoracolumbar spine: a retrospective study.
Elshamly, Mahmoud; Windhager, Reinhard; Toegel, Stefan; Grohs, Josef Georg.
Afiliación
  • Elshamly M; Department of Orthopedics and Trauma Surgery, Division of orthopedics, Medical University of Vienna, Waehringer Guertel 18-20, A-1090, Vienna, Austria.
  • Windhager R; Karl Chiari Lab for Orthopedic Biology, Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Waehringer Guertel 18-20, A-1090, Vienna, Austria.
  • Toegel S; Department of Orthopedics and Trauma Surgery, Division of orthopedics, Medical University of Vienna, Waehringer Guertel 18-20, A-1090, Vienna, Austria.
  • Grohs JG; Karl Chiari Lab for Orthopedic Biology, Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Waehringer Guertel 18-20, A-1090, Vienna, Austria.
BMC Musculoskelet Disord ; 21(1): 387, 2020 Jun 16.
Article en En | MEDLINE | ID: mdl-32546153
ABSTRACT

BACKGROUND:

The importance of sagittal alignment in healthy individuals and in reconstructive spinal surgery has been studied over the last 15 years. The aim of the present study was to assess the long-term effects of abnormal sagittal alignment on hardware after posterior thoracolumbar spinal fusion.

METHODS:

Patients who had undergone revision surgery (revision cohort, n = 34) due to breakage of their implants were compared retrospectively with patients who had intact implants at the final follow-up investigation after a long posterior thoracolumbar and/or lumbar spinal fusion (control cohort, n = 22). Clinical data and radiological parameters including the sagittal vertical axis (SVA), pelvic incidence (PI), lordosis gap (LG), pelvic tilt (PT), sacral slope (SS), lumbar lordosis (LL), thoracic kyphosis (TK), and the femoral obliquity angle (FOA) were assessed on full-spine lateral radiographs obtained in regular standing position. Data were analysed using descriptive statistics, parametric and non-parametric inferential statistics.

RESULTS:

Patients in the breakage group (female n = 21, male n = 9, mean age 60.9 ± 15.6 years) had a higher anterior shift of the C7 plumb line (SVA) (p = 0.02), retroversion of the pelvis (PT) (p < 0.001), PI-LL mismatch (LG) (p = 0.001), and PI (p = 0.002) than the intact group (female n = 10, male n = 12, mean age 65.7 ± 12.4 years). No significant difference was registered between groups in regard of SS, LL, TK, FOA, and the mean number of comorbidities.

CONCLUSION:

Failure of restoration of the SVA and the LG to the acceptable ranges, especially in patients with a high PI, may be regarded as a risk factor for the long-term failure of implants after posterior thoracolumbar spinal fusion.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Escoliosis / Fusión Vertebral / Vértebras Torácicas / Vértebras Lumbares Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: BMC Musculoskelet Disord Asunto de la revista: FISIOLOGIA / ORTOPEDIA Año: 2020 Tipo del documento: Article País de afiliación: Austria

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Escoliosis / Fusión Vertebral / Vértebras Torácicas / Vértebras Lumbares Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: BMC Musculoskelet Disord Asunto de la revista: FISIOLOGIA / ORTOPEDIA Año: 2020 Tipo del documento: Article País de afiliación: Austria