Effect of radioprotective curtain length on the scattered dose rate distribution and endoscopist eye lens dose with an over-couch fluoroscopy system.
Matsubara, Kosuke; Nakajima, Asuka; Hirosawa, Ayaka; Yoshikawa, Ryo; Ichikawa, Nao; Fukushima, Kotaro; Fukuda, Atsushi.
Afiliación
  • Matsubara K; Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan. matsuk@mhs.mp.kanazawa-u.ac.jp.
  • Nakajima A; Department of Radiological Technology, School of Health Sciences, College of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan.
  • Hirosawa A; Division of Radiology, Shinshu University Hospital, 3-1-1 Asahi, Matsumoto, Nagano, 390-8621, Japan.
  • Yoshikawa R; Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan.
  • Ichikawa N; Department of Medical Technology, Toyama Prefectural Central Hospital, 2-2-78 Nishinagae, Toyoma, Toyama, 930-8550, Japan.
  • Fukushima K; Department of Radiology, Kanazawa University Hospital, 13-1 Takaramachi, Kanazawa, Ishikawa, 920-8641, Japan.
  • Fukuda A; Department of Quantum Medical Technology, Division of Health Sciences, Graduate School of Medical Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan.
Phys Eng Sci Med ; 47(2): 691-701, 2024 Jun.
Article en En | MEDLINE | ID: mdl-38483784
ABSTRACT
Sufficient dose reduction may not be achieved if radioprotective curtains are folded. This study aimed to evaluate the scattered dose rate distribution and physician eye lens dose at different curtain lengths. Using an over-couch fluoroscopy system, dH*(10)/dt was measured using a survey meter 150 cm from the floor at 29 positions in the examination room when the curtain lengths were 0% (no curtain), 50%, 75%, and 100%. The absorbed dose rates in the air at the positions of endoscopist and assistant were calculated using a Monte Carlo simulation by varying the curtain length from 0 to 100%. The air kerma was measured by 10 min fluoroscopy using optically stimulated luminescence dosimeters at the eye surfaces of the endoscopist phantom and the outside and inside of the radioprotective goggles. At curtain lengths of 50%, 75%, and 100%, the ratios of dH*(10)/dt relative to 0% ranged from 80.8 to 104.1%, 10.5 to 61.0%, and 11.8 to 24.8%, respectively. In the simulation, the absorbed dose rates at the endoscopist's and assistant's positions changed rapidly between 55 and 75% and 65% and 80% of the curtain length, respectively. At the 0%, 50%, 75%, and 100% curtain lengths, the air kerma at the left eye surface of the endoscopist phantom was 237 ± 29, 271 ± 30, 37.7 ± 7.5, and 33.5 ± 6.1 µGy, respectively. Therefore, a curtain length of 75% or greater is required to achieve a sufficient eye lens dose reduction effect at the position of the endoscopist.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Protección Radiológica / Dispersión de Radiación / Método de Montecarlo / Cristalino Límite: Humans Idioma: En Revista: Phys Eng Sci Med Año: 2024 Tipo del documento: Article País de afiliación: Japón Pais de publicación: Suiza

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Dosis de Radiación / Protección Radiológica / Dispersión de Radiación / Método de Montecarlo / Cristalino Límite: Humans Idioma: En Revista: Phys Eng Sci Med Año: 2024 Tipo del documento: Article País de afiliación: Japón Pais de publicación: Suiza