Your browser doesn't support javascript.
loading
Side differences and reproducibility of the Moxy muscle oximeter during cycling in trained men.
Skotzke, Philip; Schwindling, Sascha; Meyer, Tim.
Afiliación
  • Skotzke P; Institute of Sport and Preventive Medicine, University of Saarland, Campus B8.2, 66123, Saarbrücken, Germany. s8phskot@uni-saarland.de.
  • Schwindling S; Institute of Sport and Preventive Medicine, University of Saarland, Campus B8.2, 66123, Saarbrücken, Germany.
  • Meyer T; Institute of Sport and Preventive Medicine, University of Saarland, Campus B8.2, 66123, Saarbrücken, Germany.
Eur J Appl Physiol ; 124(10): 3075-3083, 2024 Oct.
Article en En | MEDLINE | ID: mdl-38809481
ABSTRACT

PURPOSE:

Portable near-infrared spectroscopy devices allow measurements of muscle oxygen saturation (SmO2) in real time and non-invasively. To use NIRS for typical applications including intensity control and load monitoring, the day-to-day variability needs to be known to interpret changes confidently. This study investigates the absolute and relative test-retest reliability of the Moxy Monitor and investigates side differences of SmO2 at the vastus lateralis muscle of both legs in cyclists.

METHODS:

Twelve trained cyclists and triathletes completed 3 incremental step tests with 5 min step duration starting at 1.0 W/kg with an increase of 0.5 W/kg separated by 2-7 days. SmO2 was averaged over the last minute of each stage. For all power outputs, the intra-class coefficient (ICC), the standard error of measurement (SEM) and the minimal detectable change (MDC) were calculated. Dominant and non-dominant leg SmO2 were compared using a three-factor ANOVA and limits of agreement (LoA).

RESULTS:

ANOVA showed no significant systematic differences between trials and side. For both legs and all intensities, the ICC ranged from 0.79 to 0.92, the SEM from 5 to 9% SmO2 and the MDC from 14 to 18% SmO2. The bias and LoA between both legs were -2.0% ± 19.9% SmO2.

CONCLUSION:

Relative reliability of SmO2 was numerically good to excellent according to current standards. However, it depends on the specific analytical goal whether the test-retest reliability is deemed sufficient. Wide LoA indicate side differences in muscle oxygenation during exercise unexplained by leg dominance.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ciclismo / Oximetría Límite: Adult / Humans / Male Idioma: En Revista: Eur J Appl Physiol Asunto de la revista: FISIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Alemania Pais de publicación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Ciclismo / Oximetría Límite: Adult / Humans / Male Idioma: En Revista: Eur J Appl Physiol Asunto de la revista: FISIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Alemania Pais de publicación: Alemania