Photon counting Raman spectroscopy: a benchmarking study vs surface plasmon enhancement.
Opt Express
; 32(10): 16657-16669, 2024 May 06.
Article
en En
| MEDLINE
| ID: mdl-38858866
ABSTRACT
We demonstrate a single-photon counting Raman spectroscope and benchmark it against conventional and surface-enhanced Raman spectroscopy. For direct comparison without ambiguity, we use the same solutions of Rhodamine 6G and a common optical setup with either a spectrometer or an acousto-optic tunable filter, whereas the surface enhancement is realized with immobilized Ag nanoparticles. Our results find that the single photon counting significantly elevates the detection sensitivity by up to eight orders of magnitude, arriving at a comparable level of surface-enhanced Raman spectroscopy. Another significant advantage is with the time-resolving measurement, where we demonstrate time-gated and time-correlated single-photon counting with sub-nanosecond resolution. It offers insights into the samples' transient responses and enables the isolation of Raman scattering from fluorescence signals.
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Idioma:
En
Revista:
Opt Express
Asunto de la revista:
OFTALMOLOGIA
Año:
2024
Tipo del documento:
Article
Pais de publicación:
Estados Unidos