Comparing quantitative image parameters between animal and clinical CT-scanners: a translational phantom study analysis.
Vellala, Abhinay; Mogler, Carolin; Haag, Florian; Tollens, Fabian; Rudolf, Henning; Pietsch, Friedrich; Wängler, Carmen; Wängler, Björn; Schoenberg, Stefan O; Froelich, Matthias F; Hertel, Alexander.
Afiliación
  • Vellala A; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Mogler C; Department of Pathology, Technical University of Munich, Munich, Germany.
  • Haag F; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Tollens F; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Rudolf H; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Pietsch F; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Wängler C; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Wängler B; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Schoenberg SO; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Froelich MF; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
  • Hertel A; Department of Radiology and Nuclear medicine, University Medical Centre Mannheim, University of Heidelberg, Mannheim, Germany.
Front Med (Lausanne) ; 11: 1407235, 2024.
Article en En | MEDLINE | ID: mdl-38903806
ABSTRACT

Purpose:

This study compares phantom-based variability of extracted radiomics features from scans on a photon counting CT (PCCT) and an experimental animal PET/CT-scanner (Albira II) to investigate the potential of radiomics for translation from animal models to human scans. While oncological basic research in animal PET/CT has allowed an intrinsic comparison between PET and CT, but no 11 translation to a human CT scanner due to resolution and noise limitations, Radiomics as a statistical and thus scale-independent method can potentially close the critical gap.

Methods:

Two phantoms were scanned on a PCCT and animal PET/CT-scanner with different scan parameters and then the radiomics parameters were extracted. A Principal Component Analysis (PCA) was conducted. To overcome the limitation of a small dataset, a data augmentation technique was applied. A Ridge Classifier was trained and a Feature Importance- and Cluster analysis was performed.

Results:

PCA and Cluster Analysis shows a clear differentiation between phantom types while emphasizing the comparability of both scanners. The Ridge Classifier exhibited a strong training performance with 93% accuracy, but faced challenges in generalization with a test accuracy of 62%.

Conclusion:

These results show that radiomics has great potential as a translational tool between animal models and human routine diagnostics, especially using the novel photon counting technique. This is another crucial step towards integration of radiomics analysis into clinical practice.